Effect of microchannel geometry and linker molecules on surface immobilization efficiency of proteins in microfluidic devices.
J Biotechnol
; 364: 31-39, 2023 Feb 20.
Article
em En
| MEDLINE
| ID: mdl-36702256
One of the key metrics in the design of biosensors is the presence of an effective capture layer. Surface-immobilized proteins (especially as a part of antibody-antigen combinations) are the most commonly used capture ligands in biosensors. The surface coverage of these proteins in flow-based biosensors are affected by both the linker chemistry used to attach them as well as the microchannel geometry. We used streptavidin as a model protein to compare glutaraldehyde, EDC-NHS, sulfo-SMCC and sulfo-NHS-biotin as linkers inside straight, serpentine and square-wave microchannel geometries. We found that straight microchannels achieve the highest degree of protein immobilization compared to serpentine and square-wave microchannels, irrespective of the linker chemistry used. We also showed that for a given microchannel geometry, sulfo-NHS-biotin leads to the highest immobilization of streptavidin among all the linkers.