Bioimpedance Analysis as Early Predictor for Clot Formation Inside a Blood-Perfused Test Chamber: Proof of Concept Using an In Vitro Test-Circuit.
Türkmen, Muhammet; Lauwigi, Tobias; Fechter, Tamara; Gries, Fabienne; Fischbach, Anna; Gries, Thomas; Rossaint, Rolf; Bleilevens, Christian; Winnersbach, Patrick.
Afiliação
  • Türkmen M; Department of Anesthesiology, University Hospital RWTH Aachen, Pauwelsstraße 30, 52074 Aachen, Germany.
  • Lauwigi T; Institut für Textiltechnik (ITA), RWTH Aachen University, 52074 Aachen, Germany.
  • Fechter T; Department of Anesthesiology, University Hospital RWTH Aachen, Pauwelsstraße 30, 52074 Aachen, Germany.
  • Gries F; Institut für Textiltechnik (ITA), RWTH Aachen University, 52074 Aachen, Germany.
  • Fischbach A; Department of Anesthesiology, University Hospital RWTH Aachen, Pauwelsstraße 30, 52074 Aachen, Germany.
  • Gries T; Institut für Textiltechnik (ITA), RWTH Aachen University, 52074 Aachen, Germany.
  • Rossaint R; Department of Anesthesiology, University Hospital RWTH Aachen, Pauwelsstraße 30, 52074 Aachen, Germany.
  • Bleilevens C; Department of Anesthesiology, University Hospital RWTH Aachen, Pauwelsstraße 30, 52074 Aachen, Germany.
  • Winnersbach P; Department of Anesthesiology, University Hospital RWTH Aachen, Pauwelsstraße 30, 52074 Aachen, Germany.
Biosensors (Basel) ; 13(3)2023 Mar 17.
Article em En | MEDLINE | ID: mdl-36979606
Clot formation inside a membrane oxygenator (MO) due to blood-to-foreign surface interaction represents a frequent complication during extracorporeal membrane oxygenation. Since current standard monitoring methods of coagulation status inside the MO fail to detect clot formation at an early stage, reliable sensors for early clot detection are in demand to reduce associated complications and adverse events. Bioimpedance analysis offers a monitoring concept by integrating sensor fibers into the MO. Herein, the feasibility of clot detection via bioimpedance analysis is evaluated. A custom-made test chamber with integrated titanium fibers acting as sensors was perfused with heparinized human whole blood in an in vitro test circuit until clot formation occurred. The clot detection capability of bioimpedance analysis was directly compared to the pressure difference across the test chamber (ΔP-TC), analogous to the measurement across MOs (ΔP-MO), the clinical gold standard for clot detection. We found that bioimpedance measurement increased significantly 8 min prior to a significant increase in ΔP-TC, indicating fulminant clot formation. Experiments without clot formation resulted in a lack of increase in bioimpedance or ΔP-TC. This study shows that clot detection via bioimpedance analysis under flow conditions in a blood-perfused test chamber is generally feasible, thus paving the way for further investigation.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Trombose / Oxigenação por Membrana Extracorpórea Tipo de estudo: Diagnostic_studies / Etiology_studies / Prognostic_studies / Risk_factors_studies Limite: Humans Idioma: En Revista: Biosensors (Basel) Ano de publicação: 2023 Tipo de documento: Article País de afiliação: Alemanha País de publicação: Suíça

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Trombose / Oxigenação por Membrana Extracorpórea Tipo de estudo: Diagnostic_studies / Etiology_studies / Prognostic_studies / Risk_factors_studies Limite: Humans Idioma: En Revista: Biosensors (Basel) Ano de publicação: 2023 Tipo de documento: Article País de afiliação: Alemanha País de publicação: Suíça