Your browser doesn't support javascript.
loading
Intraoperative Beat-to-Beat Pulse Transit Time (PTT) Monitoring via Non-Invasive Piezoelectric/Piezocapacitive Peripheral Sensors Can Predict Changes in Invasively Acquired Blood Pressure in High-Risk Surgical Patients.
Nordine, Michael; Pille, Marius; Kraemer, Jan; Berger, Christian; Brandhorst, Philipp; Kaeferstein, Philipp; Kopetsch, Roland; Wessel, Niels; Trauzeddel, Ralf Felix; Treskatsch, Sascha.
Afiliação
  • Nordine M; Department of Anesthesiology and Intensive Care Medicine, Hindenburgdamm 30, Charité-Universitätsmedizin Berlin, Corporate Member of Freie Universität Berlin and Humboldt Universität zu Berlin, 12203 Berlin, Germany.
  • Pille M; Berlin Institute of Health at Charité, Universitätsmedizin Berlin, Charitéplatz 1, 10117 Berlin, Germany.
  • Kraemer J; Department of Physics, Humboldt University zu Berlin, 10115 Berlin, Germany.
  • Berger C; Department of Physics, Humboldt University zu Berlin, 10115 Berlin, Germany.
  • Brandhorst P; Department of Anesthesiology and Intensive Care Medicine, Hindenburgdamm 30, Charité-Universitätsmedizin Berlin, Corporate Member of Freie Universität Berlin and Humboldt Universität zu Berlin, 12203 Berlin, Germany.
  • Kaeferstein P; Department of Anesthesiology and Intensive Care Medicine, Hindenburgdamm 30, Charité-Universitätsmedizin Berlin, Corporate Member of Freie Universität Berlin and Humboldt Universität zu Berlin, 12203 Berlin, Germany.
  • Kopetsch R; SectorCon-Ingenieurgesellschaft mbH, 10553 Berlin, Germany.
  • Wessel N; SectorCon-Ingenieurgesellschaft mbH, 10553 Berlin, Germany.
  • Trauzeddel RF; Department of Physics, Humboldt University zu Berlin, 10115 Berlin, Germany.
  • Treskatsch S; Department of Human Medicine, MSB Medical School Berlin GmbH, 14197 Berlin, Germany.
Sensors (Basel) ; 23(6)2023 Mar 21.
Article em En | MEDLINE | ID: mdl-36992016
ABSTRACT

BACKGROUND:

Non-invasive tracking of beat-to-beat pulse transit time (PTT) via piezoelectric/piezocapacitive sensors (PES/PCS) may expand perioperative hemodynamic monitoring. This study evaluated the ability for PTT via PES/PCS to correlate with systolic, diastolic, and mean invasive blood pressure (SBPIBP, DBPIBP, and MAPIBP, respectively) and to detect SBPIBP fluctuations.

METHODS:

PES/PCS and IBP measurements were performed in 20 patients undergoing abdominal, urological, and cardiac surgery. A Pearson's correlation analysis (r) between 1/PTT and IBP was performed. The predictive ability of 1/PTT with changes in SBPIBP was determined by area under the curve (reported as AUC, sensitivity, specificity).

RESULTS:

Significant correlations between 1/PTT and SBPIBP were found for PES (r = 0.64) and PCS (r = 0.55) (p < 0.01), as well as MAPIBP/DBPIBP for PES (r = 0.6/0.55) and PCS (r = 0.5/0.45) (p < 0.05). A 7% decrease in 1/PTTPES predicted a 30% SBPIBP decrease (0.82, 0.76, 0.76), while a 5.6% increase predicted a 30% SBPIBP increase (0.75, 0.7, 0.68). A 6.6% decrease in 1/PTTPCS detected a 30% SBPIBP decrease (0.81, 0.72, 0.8), while a 4.8% 1/PTTPCS increase detected a 30% SBPIBP increase (0.73, 0.64, 0.68).

CONCLUSIONS:

Non-invasive beat-to-beat PTT via PES/PCS demonstrated significant correlations with IBP and detected significant changes in SBPIBP. Thus, PES/PCS as a novel sensor technology may augment intraoperative hemodynamic monitoring during major surgery.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Monitorização Intraoperatória / Pressão Arterial Tipo de estudo: Etiology_studies / Prognostic_studies / Risk_factors_studies Limite: Humans Idioma: En Revista: Sensors (Basel) Ano de publicação: 2023 Tipo de documento: Article País de afiliação: Alemanha

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Monitorização Intraoperatória / Pressão Arterial Tipo de estudo: Etiology_studies / Prognostic_studies / Risk_factors_studies Limite: Humans Idioma: En Revista: Sensors (Basel) Ano de publicação: 2023 Tipo de documento: Article País de afiliação: Alemanha