Your browser doesn't support javascript.
loading
Is an Anatomical Suprapectineal Quadrilateral Surface Plate Superior to Previous Fixation Methods for Anterior Column-Posterior Hemitransverse Acetabular Fractures Typical in the Elderly?: A Biomechanical Study.
Kwak, Dae-Kyung; Jang, Ji-Eun; Kim, Won-Hyeon; Lee, Sung-Jae; Lee, Yongmin; Yoo, Je-Hyun.
Afiliação
  • Kwak DK; Department of Orthopaedic Surgery, Hallym University Sacred Heart Hospital, Hallym University School of Medicine, Anyang, Korea.
  • Jang JE; Department of Biomedical Engineering, Inje University, Gimhae, Korea.
  • Kim WH; Department of Mechanical Engineering, Sejong University, Seoul, Korea.
  • Lee SJ; Department of Biomedical Engineering, Inje University, Gimhae, Korea.
  • Lee Y; Department of Orthopaedic Surgery, Hallym University Sacred Heart Hospital, Hallym University School of Medicine, Anyang, Korea.
  • Yoo JH; Department of Orthopaedic Surgery, Hallym University Sacred Heart Hospital, Hallym University School of Medicine, Anyang, Korea.
Clin Orthop Surg ; 15(2): 182-191, 2023 Apr.
Article em En | MEDLINE | ID: mdl-37008963
ABSTRACT

Background:

This study aimed to compare the mechanical characteristics of four fixation methods including an anatomical suprapectineal quadrilateral surface (QLS) plate in hemipelvic models of anterior column-posterior hemitransverse acetabular fractures typical in elderly patients.

Methods:

In total, 24 composite hemipelvic models were used and allocated to four groups group 1, pre-contoured anatomical suprapectineal QLS plate; group 2, suprapectineal reconstruction plate with two periarticular long screws; group 3, suprapectineal reconstruction plate with a buttress reconstruction plate; group 4, suprapectineal reconstruction plate with a buttress T-plate. Axial structural stiffness and displacement of each column fragment in four different fixation constructs were compared.

Results:

Multiple group comparisons of axial structural stiffness demonstrated significant difference (p = 0.001). Although there was no significant difference between groups 1 and 2 (p = 0.699), group 1 showed greater stiffness than groups 3 and 4 (p = 0.002 and 0.002, respectively). Group 1 showed less displacement in the anterior region of the anterior fragment than group 4 (p = 0.009) and in the posterior region than groups 3 and 4 (p = 0.015 and p = 0.015, respectively). However, group 1 demonstrated greater displacement than group 2 in the posterior region of the posterior fragment (p = 0.004), while showing similar displacement to groups 3 and 4.

Conclusions:

The anatomical suprapectineal QLS plate provided the mechanical stability comparable or superior to other existing fixations in osteoporotic models of anterior column-posterior hemitransverse acetabular fractures typical in the elderly. However, additional plate modification would be needed for better stability and outcomes.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Fraturas da Coluna Vertebral / Fraturas Ósseas / Fraturas do Quadril Tipo de estudo: Prognostic_studies Limite: Aged / Humans Idioma: En Revista: Clin Orthop Surg Assunto da revista: ORTOPEDIA Ano de publicação: 2023 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Fraturas da Coluna Vertebral / Fraturas Ósseas / Fraturas do Quadril Tipo de estudo: Prognostic_studies Limite: Aged / Humans Idioma: En Revista: Clin Orthop Surg Assunto da revista: ORTOPEDIA Ano de publicação: 2023 Tipo de documento: Article