Your browser doesn't support javascript.
loading
Multi-vendor robustness analysis of a commercial artificial intelligence system for breast cancer detection.
Riveira-Martin, Mercedes; Rodríguez-Ruiz, Alejandro; Martí, Robert; Chevalier, Margarita.
Afiliação
  • Riveira-Martin M; Complutense University of Madrid, Medicine School, Department of Radiology, Rehabilitation and Physiotherapy, Madrid, Spain.
  • Rodríguez-Ruiz A; Galicia Sur Health Research Institute, Meixoeiro Hospital, Department of Medical Physics and RP, Vigo, Spain.
  • Martí R; Nicolab BV, Amsterdam, The Netherlands.
  • Chevalier M; University of Girona, Computer Vision and Robotics Institute (VICOROB), Girona, Spain.
J Med Imaging (Bellingham) ; 10(5): 051807, 2023 Sep.
Article em En | MEDLINE | ID: mdl-37082509
ABSTRACT

Purpose:

Population-based screening programs for the early detection of breast cancer have significantly reduced mortality in women, but they are resource intensive in terms of time, cost, and workload and still have limitations mainly due to the use of 2D imaging techniques, which may cause overlapping of tissues, and interobserver variability. Artificial intelligence (AI) systems may be a valuable tool to assist radiologist when reading and classifying mammograms based on the malignancy of the detected lesions. However, there are several factors that can influence the outcome of a mammogram and thus also the detection capability of an AI system. The aim of our work is to analyze the robustness of the diagnostic ability of an AI system designed for breast cancer detection.

Approach:

Mammograms from a population-based screening program were scored with the AI system. The sensitivity and specificity by means of the area under the receiver operating characteristic (ROC) curve were obtained as a function of the mammography unit manufacturer, demographic characteristics, and several factors that may affect the image quality (age, breast thickness and density, compression applied, beam quality, and delivered dose).

Results:

The area under the curve (AUC) from the scoring ROC curve was 0.92 (95% confidence interval = 0.89 - 0.95). It showed no dependence with any of the parameters considered, as the differences in the AUC for different interval values were not statistically significant.

Conclusion:

The results suggest that the AI system analyzed in our work has a robust diagnostic capability, and that its accuracy is independent of the studied parameters.
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Tipo de estudo: Diagnostic_studies / Prognostic_studies / Screening_studies Idioma: En Revista: J Med Imaging (Bellingham) Ano de publicação: 2023 Tipo de documento: Article País de afiliação: Espanha País de publicação: EEUU / ESTADOS UNIDOS / ESTADOS UNIDOS DA AMERICA / EUA / UNITED STATES / UNITED STATES OF AMERICA / US / USA

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Tipo de estudo: Diagnostic_studies / Prognostic_studies / Screening_studies Idioma: En Revista: J Med Imaging (Bellingham) Ano de publicação: 2023 Tipo de documento: Article País de afiliação: Espanha País de publicação: EEUU / ESTADOS UNIDOS / ESTADOS UNIDOS DA AMERICA / EUA / UNITED STATES / UNITED STATES OF AMERICA / US / USA