Optimizing variable flip angles in magnetization-prepared gradient-echo sequences for efficient 3D-T1ρ mapping.
Zibetti, Marcelo V W; De Moura, Hector L; Keerthivasan, Mahesh B; Regatte, Ravinder R.
Afiliação
  • Zibetti MVW; Center for Biomedical Imaging, Department of Radiology, New York University Grossman School of Medicine, New York, New York, USA.
  • De Moura HL; Center for Biomedical Imaging, Department of Radiology, New York University Grossman School of Medicine, New York, New York, USA.
  • Keerthivasan MB; Siemens Medical Solutions USA, Malvern, Pennsylvania, USA.
  • Regatte RR; Center for Biomedical Imaging, Department of Radiology, New York University Grossman School of Medicine, New York, New York, USA.
Magn Reson Med ; 90(4): 1465-1483, 2023 10.
Article em En | MEDLINE | ID: mdl-37288538
PURPOSE: To optimize the choice of the flip angles of magnetization-prepared gradient-echo sequences for improved accuracy, precision, and speed of 3D-T1ρ mapping. METHODS: We propose a new optimization approach for finding variable flip-angle values that improve magnetization-prepared gradient-echo sequences used for 3D-T1ρ mapping. This new approach can improve the accuracy and SNR, while reducing filtering effects. We demonstrate the concept in the three different versions of the magnetization-prepared gradient-echo sequences that are typically used for 3D-T1ρ mapping and evaluate their performance in model agarose phantoms (n = 4) and healthy volunteers (n = 5) for knee joint imaging. We also tested the optimization with sequence parameters targeting faster acquisitions. RESULTS: Our results show that optimized variable flip angle can improve the accuracy and the precision of the sequences, seen as a reduction of the mean of normalized absolute difference from about 5%-6% to 3%-4% in model phantoms and from 15%-16% to 11%-13% in the knee joint, and improving SNR from about 12-28 to 22-32 in agarose phantoms and about 7-14 to 13-17 in healthy volunteers. The optimization can also compensate for the loss in quality caused by making the sequence faster. This results in sequence configurations that acquire more data per unit of time with SNR and mean of normalized absolute difference measurements close to its slower versions. CONCLUSION: The optimization of the variable flip angle can be used to increase accuracy and precision, and to improve the speed of the typical imaging sequences used for quantitative 3D-T1ρ mapping of the knee joint.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Imageamento por Ressonância Magnética / Imageamento Tridimensional Tipo de estudo: Prognostic_studies Limite: Humans Idioma: En Revista: Magn Reson Med Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2023 Tipo de documento: Article País de afiliação: Estados Unidos País de publicação: Estados Unidos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Imageamento por Ressonância Magnética / Imageamento Tridimensional Tipo de estudo: Prognostic_studies Limite: Humans Idioma: En Revista: Magn Reson Med Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2023 Tipo de documento: Article País de afiliação: Estados Unidos País de publicação: Estados Unidos