Reducing contrast-agent volume and radiation dose in CT with 90-kVp tube voltage, high tube current modulation, and advanced iteration algorithm.
Park, Min Su; Ha, Hong Il; Ahn, Jhii-Hyun; Lee, In Jae; Lim, Hyun Kyung.
Afiliação
  • Park MS; Department of Radiology, Hallym University Sacred Heart Hospital, Anyang-si, Gyeonggi-do, Republic of Korea.
  • Ha HI; Department of Radiology, Hallym University Sacred Heart Hospital, Anyang-si, Gyeonggi-do, Republic of Korea.
  • Ahn JH; Department of Radiology, Yonsei University Wonju College of Medicine, Wonju Severance Christian Hospital, Wonju, Gangwon-do, Republic of Korea.
  • Lee IJ; Department of Radiology, Hallym University Sacred Heart Hospital, Anyang-si, Gyeonggi-do, Republic of Korea.
  • Lim HK; Department of Radiology, Soonchunhyang University Seoul Hospital, Seoul, Republic of Korea.
PLoS One ; 18(6): e0287214, 2023.
Article em En | MEDLINE | ID: mdl-37319309
ABSTRACT
Increasing utilization of computed tomography (CT) has raised concerns regarding CT radiation dose and technology has been developed to achieve an appropriate balance between image quality, radiation dose, and the amount of contrast material. This study was planned to evaluate the image quality and radiation dose in pancreatic dynamic computed tomography (PDCT) with 90-kVp tube voltage and reduction of the standard amount of contrast agent, compared with 100-kVp PDCT of the research hospital's convention. Total of 51 patients with both CT protocols were included. The average Hounsfield units (HU) values of the abdominal organs and image noise were measured for objective image quality analysis. Two radiologists evaluated five categories of image qualities such as subjective image noise, visibility of small structure, beam hardening or streak artifact, lesion conspicuity and overall diagnostic performance for subjective image quality analysis. The total amount of contrast agent, radiation dose, and image noise decreased in the low-kVp group, by 24.4%, 31.7%, and 20.6%, respectively (p < 0.001). The intraobserver and interobserver agreements were moderate to substantial (k = 0.4-0.8). The contrast-to-noise ratio (CNR), signal-to-noise ratio (SNR), and figure of merit of the almost organs except psoas muscle in the low-kVp group were significantly higher (p < 0.001). Except for lesion conspicuity, both reviewers judged that subjective image quality of the 90-kVp group was better (p < 0.001). With 90-kVp tube voltage, 25% reduced contrast agent volume with advanced iteration algorithm and high tube current modulation achieved radiation dose reduction of 31.7%, as well as better image quality and diagnostic confidence.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Tomografia Computadorizada por Raios X / Meios de Contraste Tipo de estudo: Guideline / Prognostic_studies Limite: Humans Idioma: En Revista: PLoS One Assunto da revista: CIENCIA / MEDICINA Ano de publicação: 2023 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Tomografia Computadorizada por Raios X / Meios de Contraste Tipo de estudo: Guideline / Prognostic_studies Limite: Humans Idioma: En Revista: PLoS One Assunto da revista: CIENCIA / MEDICINA Ano de publicação: 2023 Tipo de documento: Article