Newly Designed 3D-Printed Sonication Test Cell Optimized for In Vitro Sonication Experiments.
Snehota, Martin; Kolarikova, Marketa; Vachutka, Jaromir; Hosikova, Barbora; Balazova, Klara; Dolezal, Ladislav; Kolarova, Hana.
Afiliação
  • Snehota M; Department of Medical Biophysics, Faculty of Medicine and Dentistry, Palacky University Olomouc, Olomouc, Czech Republic.
  • Kolarikova M; Department of Medical Biophysics, Faculty of Medicine and Dentistry, Palacky University Olomouc, Olomouc, Czech Republic.
  • Vachutka J; Department of Medical Biophysics, Faculty of Medicine and Dentistry, Palacky University Olomouc, Olomouc, Czech Republic.
  • Hosikova B; Department of Medical Biophysics, Faculty of Medicine and Dentistry, Palacky University Olomouc, Olomouc, Czech Republic. Electronic address: barbora.hosikova@upol.cz.
  • Balazova K; Department of Medical Biophysics, Faculty of Medicine and Dentistry, Palacky University Olomouc, Olomouc, Czech Republic.
  • Dolezal L; Department of Medical Biophysics, Faculty of Medicine and Dentistry, Palacky University Olomouc, Olomouc, Czech Republic.
  • Kolarova H; Department of Medical Biophysics, Faculty of Medicine and Dentistry, Palacky University Olomouc, Olomouc, Czech Republic.
Ultrasound Med Biol ; 49(9): 2042-2053, 2023 09.
Article em En | MEDLINE | ID: mdl-37330389
ABSTRACT

OBJECTIVE:

Precise control over the ultrasound field parameters experienced by biological samples during sonication experiments in vitro may be quite challenging. The main goal of this work was to outline an approach to construction of sonication test cells that would minimize the interaction between the test cells and ultrasound.

METHODS:

Optimal dimensions of the test cell were determined through measurements conducted in a water sonication tank using 3D-printed test objects. The offset of local acoustic intensity variability inside the sonication test cell was set to value of ±50% of the reference value (i.e., local acoustic intensity measured at last axial maximum in the free-field condition). The cytotoxicity of several materials used for 3D printing was determined using the MTT (3-[4,5-dimethylthiazol-2-yl]-2,5 diphenyl tetrazolium bromide) assay.

RESULTS:

The sonication test cells were 3D printed from polylactic acid material, which was not toxic to the cells. Silicone membrane HT-6240, which was used to construct the bottom of the test cell, was found to reduce ultrasound energy minimally. Final ultrasound profiles inside the sonication test cells indicated the desired variability of local acoustic intensity. The cell viability in our sonication test cell was comparable to that of commercial culture plates with bottoms constructed with silicone membrane.

CONCLUSION:

An approach to construction of sonication test cells minimizing the interaction of the test cell and ultrasound has been outlined.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Sonicação / Ablação por Ultrassom Focalizado de Alta Intensidade Tipo de estudo: Diagnostic_studies Idioma: En Revista: Ultrasound Med Biol Ano de publicação: 2023 Tipo de documento: Article País de afiliação: República Tcheca

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Sonicação / Ablação por Ultrassom Focalizado de Alta Intensidade Tipo de estudo: Diagnostic_studies Idioma: En Revista: Ultrasound Med Biol Ano de publicação: 2023 Tipo de documento: Article País de afiliação: República Tcheca