Analytical validation, sample stability, and clinical evaluation of a new high-sensitivity cardiac troponin I immunoassay for use in dogs, with comparison to a previous ultrasensitive assay.
PLoS One
; 18(7): e0288801, 2023.
Article
em En
| MEDLINE
| ID: mdl-37463140
Cardiac troponin I (cTnI) is considered the gold standard biomarker for myocardial injury and shows a high degree of homology between humans and dogs. The ADVIA Centaur XP High-Sensitivity Troponin I (AC-cTnI-HS) assay has been validated for use in humans but not dogs. The study objectives were to analytically validate the AC-cTnI-HS assay in dogs, to assess correlation between the AC-cTnI-HS and a previous ADVIA Centaur TnI-Ultra (AC-cTnI-U) assay, to assess cTnI sample storage stability, and to clinically evaluate the AC-cTnI-HS assay in healthy dogs and dogs with cardiac disease. Canine serum samples were used for analytical validation. Intra- and inter-assay variability, dilutional parallelism, and spiking recovery were assessed. Samples from 196 client-owned dogs were evaluated (healthy dogs (n = 39) or dogs with congenital heart disease (n = 54), myxomatous mitral valve disease (n = 68), dilated cardiomyopathy (n = 15), or myocarditis (n = 20)). Inter- and intra-assay coefficient of variation (%CV) was between 2.8-41.4% and 3.8-30.2%, respectively, with pools with concentrations >20 pg/mL all having %CVs <10%. The observed to expected ratios for dilutional parallelism and spiking recovery experiments ranged between 92.3 and 266.7.0% and 84.3 and 108%, respectively. A strong correlation between the AC-cTnI-HS and AC-cTnI-U assays was observed (Spearman's ρ = 0.927), though a proportional bias existed, with AC-cTnI-HS assay concentrations being proportionally lower than AC-cTnI-U assay concentrations. Serum samples stored at -80°C had stable cTnI measurements for up to 2.7 years and after a single freeze-thaw cycle. Healthy dogs and dogs with congenital heart disease had significantly lower cTnI concentrations than dogs in the other three groups. The AC-cTnI-HS assay precisely, reproducibly, and accurately measures cTnI concentrations in dog serum with cTnI concentrations >20 pg/mL.
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Cardiomiopatia Dilatada
/
Cardiopatias
/
Doenças das Valvas Cardíacas
Tipo de estudo:
Diagnostic_studies
Limite:
Animals
/
Humans
Idioma:
En
Revista:
PLoS One
Assunto da revista:
CIENCIA
/
MEDICINA
Ano de publicação:
2023
Tipo de documento:
Article
País de afiliação:
Estados Unidos
País de publicação:
Estados Unidos