Highly sensitive and label-free detection of biotin using a liquid crystal-based optofluidic biosensor.
Biomed Opt Express
; 14(7): 3763-3774, 2023 Jul 01.
Article
em En
| MEDLINE
| ID: mdl-37497519
A liquid crystal (LC)-based optofluidic whispering gallery mode (WGM) resonator has been applied as a biosensor to detect biotin. Immobilized streptavidin (SA) act as protein molecules and specifically bind to biotin through strong non-covalent interaction, which can interfere with the orientation of LCs by decreasing the vertical anchoring force of the alignment layer in which the WGM spectral wavelength shift is monitored as a sensing parameter. Due to the double magnification of the LC molecular orientation transition and the resonance of the WGM, the detection limit for SA can reach 1.25 fM (4.7 × 10-13 g/ml). The measurable concentration of biotin and the wavelength shift of the WGM spectrum have an excellent linearity in the range of 0 to 0.1 pg/ml, which can achieve ultra-low detection limit (0.4 fM), i.e., seven orders of magnitude improvement over conventional polarized optical microscope (POM) method. The proposed optofluidic biosensor is highly reproducible and can be used as an ultrasensitive real-time monitoring biosensor, which will open the door for applications to other receptor and ligand models.
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Tipo de estudo:
Diagnostic_studies
Idioma:
En
Revista:
Biomed Opt Express
Ano de publicação:
2023
Tipo de documento:
Article
País de afiliação:
China
País de publicação: