Your browser doesn't support javascript.
loading
Investigating the Small Pixel Effect in Ultra-High Resolution Photon-Counting CT of the Lung.
Huflage, Henner; Hendel, Robin; Kunz, Andreas Steven; Ergün, Süleyman; Afat, Saif; Petri, Nils; Hartung, Viktor; Gruschwitz, Philipp; Bley, Thorsten Alexander; Grunz, Jan-Peter.
Afiliação
  • Huflage H; From the Department of Diagnostic and Interventional Radiology, University Hospital Würzburg, Würzburg, Germany (H.H., R.H., A.S., V.H., P.G., T.A., J.-P.G.); Institute of Anatomy and Cell Biology, University of Würzburg, Würzburg, Germany (S.E.); Department of Diagnostic and Interventional Radiology, University Hospital Tübingen, Tübingen, Germany (S.A.); and Department of Internal Medicine I, University Hospital Würzburg, Würzburg, Germany (N.P.).
Invest Radiol ; 59(4): 293-297, 2024 Apr 01.
Article em En | MEDLINE | ID: mdl-37552040
ABSTRACT

OBJECTIVES:

The aim of this study was to investigate potential benefits of ultra-high resolution (UHR) over standard resolution scan mode in ultra-low dose photon-counting detector CT (PCD-CT) of the lung. MATERIALS AND

METHODS:

Six cadaveric specimens were examined with 5 dose settings using tin prefiltration, each in UHR (120 × 0.2 mm) and standard mode (144 × 0.4 mm), on a first-generation PCD-CT scanner. Image quality was evaluated quantitatively by noise comparisons in the trachea and both main bronchi. In addition, 16 readers (14 radiologists and 2 internal medicine physicians) independently completed a browser-based pairwise forced-choice comparison task for assessment of subjective image quality. The Kendall rank coefficient ( W ) was calculated to assess interrater agreement, and Pearson's correlation coefficient ( r ) was used to analyze the relationship between noise measurements and image quality rankings.

RESULTS:

Across all dose levels, image noise in UHR mode was lower than in standard mode for scan protocols matched by CTDI vol ( P < 0.001). UHR examinations exhibited noise levels comparable to the next higher dose setting in standard mode ( P ≥ 0.275). Subjective ranking of protocols based on 5760 pairwise tests showed high interrater agreement ( W = 0.99; P ≤ 0.001) with UHR images being preferred by readers in the majority of comparisons. Irrespective of scan mode, a substantial indirect correlation was observed between image noise and subjective image quality ranking ( r = -0.97; P ≤ 0.001).

CONCLUSIONS:

In PCD-CT of the lung, UHR scan mode reduces image noise considerably over standard resolution acquisition. Originating from the smaller detector element size in fan direction, the small pixel effect allows for superior image quality in ultra-low dose examinations with considerable potential for radiation dose reduction.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Tomografia Computadorizada por Raios X / Fótons Limite: Humans Idioma: En Revista: Invest Radiol Ano de publicação: 2024 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Tomografia Computadorizada por Raios X / Fótons Limite: Humans Idioma: En Revista: Invest Radiol Ano de publicação: 2024 Tipo de documento: Article