Your browser doesn't support javascript.
loading
Cervical Disc Arthroplasties Fail to Maintain Physiological Kinematics Under Lateral Eccentric Loads.
Roch, Paul Jonathan; Hemprich, Constantin; Klockner, Friederike; Meier, Marc-Pascal; Jäckle, Katharina; Lehmann, Wolfgang; Wachowski, Martin Michael; Weiser, Lukas.
Afiliação
  • Roch PJ; Department of Trauma Surgery, Orthopaedics and Plastic Surgery, University of Göttingen, Göttingen, Germany.
  • Hemprich C; Department of Trauma Surgery, Orthopaedics and Plastic Surgery, University of Göttingen, Göttingen, Germany.
  • Klockner F; Department of Trauma Surgery, Orthopaedics and Plastic Surgery, University of Göttingen, Göttingen, Germany.
  • Meier MP; Department of Trauma Surgery, Orthopaedics and Plastic Surgery, University of Göttingen, Göttingen, Germany.
  • Jäckle K; Department of Trauma Surgery, Orthopaedics and Plastic Surgery, University of Göttingen, Göttingen, Germany.
  • Lehmann W; Department of Trauma Surgery, Orthopaedics and Plastic Surgery, University of Göttingen, Göttingen, Germany.
  • Wachowski MM; Department of Trauma Surgery, Orthopaedics and Plastic Surgery, University of Göttingen, Göttingen, Germany.
  • Weiser L; DUO - Duderstadt Trauma Surgery and Orthopaedics, Duderstadt, Germany.
Global Spine J ; : 21925682231193642, 2023 Aug 17.
Article em En | MEDLINE | ID: mdl-37592374
STUDY DESIGN: In vitro human cadaveric biomechanical analysis. OBJECTIVES: Optimization of prostheses for cervical disc arthroplasties (CDA) reduces the risk of complications. The instantaneous helical axis (IHA) is a superior parameter for examining the kinematics of functional spinal units. There is no comprehensive study about the IHA after CDA considering all 3 motion dimensions. METHODS: Ten human functional spinal units C4-5 (83.2 ± 7.9 yrs.) were examined with an established measuring apparatus in intact conditions (IC), and after CDA, with 2 different types of prostheses during axial rotation, lateral bending, and flexion/extension. Eccentric preloads simulated strains. The IHA orientation and its position at the point of rest (IHA0-position) were analyzed. RESULTS: The results confirmed the existing data for IHA in IC. Lateral preloads showed structural alterations of kinematics after CDA: During axial rotation and lateral bending, the shift of the IHA0-position was corresponding with the lateral preloads' applied site in IC, while after CDAs, it was vice versa. During lateral bending, the lateral IHA orientation was inclined, corresponding with the lateral preloads' applied site in the IC and oppositely after the CDAs. During flexion/extension, the lateral IHA orientation was nearly vertical in the IC, while after CDA, it inclined, corresponding with the lateral preloads' applied site. The axial IHA orientation rotated to the lateral preloads' corresponding site in the IC; after CDA, it was vice versa. CONCLUSION: Both CDAs failed to maintain physiological IHA characteristics under lateral preloads, revealing a new aspect for improving prostheses' design and optimizing their kinematics.
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Revista: Global Spine J Ano de publicação: 2023 Tipo de documento: Article País de afiliação: Alemanha País de publicação: Reino Unido