T-Piece resuscitator versus self-inflating bag for delivery room resuscitation in preterm neonates: a randomized controlled trial.
Eur J Pediatr
; 182(12): 5565-5576, 2023 Dec.
Article
em En
| MEDLINE
| ID: mdl-37792092
The establishment of adequate ventilation is the cornerstone of neonatal resuscitation in the delivery room (DR). This parallel-group, accessor-blinded randomized controlled trial compared the changes in peripheral oxygen saturation (SpO2), heart rate (HR), and cerebral regional oxygen saturation (crSO2) with the use of a T-piece resuscitator (TPR) versus self-inflating bag (SIB) as a mode of providing positive pressure ventilation (PPV) during DR resuscitation in preterm neonates. Seventy-two preterm neonates were randomly allocated to receive PPV with TPR (n = 36) or SIB (n = 36). The primary outcome was SpO2 (%) at 5 min. The secondary outcomes included the time to achieve a SpO2 ≥ 80% and > 85%, HR > 100/min, fractional-inspired oxygen (FiO2) requirement, minute-specific SpO2, HR and FiO2 trends for the first 5 min of life, need for DR-intubation, crSO2, need and duration of respiratory support, and other in-hospital morbidities. Mean SpO2 at 5 min was 74.5 ± 17.8% and 69.4 ± 22.4%, in TPR and SIB groups, respectively [Mean difference, 95% Confidence Interval 5.08 (-4.41, 14.58); p = 0.289]. No difference was observed in the time to achieve a SpO2 ≥ 80% and > 85%, HR > 100/min, the requirement of FiO2, DR-intubation, and the need and duration of respiratory support. There was no significant difference in the minute-specific SpO2, HR, and FiO2 requirements for the first 5 min. CrSO2 (%) at one hour was lower by 5% in the TPR group compared to SIB; p = 0.03. Other complications were comparable. CONCLUSIONS: TPR and SIB resulted in comparable SpO2 at 5 min along with similar minute-specific SpO2, HR, and FiO2 trends. CLINICAL TRIAL REGISTRATION: Clinical trial registry of India, Registration no: CTRI/2021/10/037384, Registered prospectively on: 20/10/2021, https://ctri.icmr.org.in/ . WHAT IS KNOWN: ⢠Compared to self-inflating bags (SIB), T-piece resuscitators (TPR) provide more consistent inflation pressure and tidal volume as shown in animal and bench studies. ⢠There is no strong recommendation for one device over the other in view of low certainty evidence. WHAT IS NEW: ⢠TPR and SIB resulted in comparable peripheral oxygen saturation (SpO2) at 5 min along with similar minute-specific SpO2, heart rate, and fractional-inspired oxygen requirement trends. ⢠Short-term complications and mortality rates were comparable with both devices.
Palavras-chave
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Ressuscitação
/
Salas de Parto
Tipo de estudo:
Clinical_trials
/
Guideline
Limite:
Humans
/
Newborn
Idioma:
En
Revista:
Eur J Pediatr
Ano de publicação:
2023
Tipo de documento:
Article
País de afiliação:
Índia
País de publicação:
Alemanha