Quantification and purification of isotopic contamination at the ReAccelerator of the Facility for Rare Isotope Beams.
Rev Sci Instrum
; 94(10)2023 Oct 01.
Article
em En
| MEDLINE
| ID: mdl-37815423
At the ReAccelerator within the Facility for Rare Isotope Beams, a combination of an interchangeable aluminum foil and a silicon detector was developed to quantify isobaric contamination in rare isotope beams. The device is simple to operate and is now used routinely. In this article, we describe the system and show an application of the device to determine the level of contamination of an Si-32 rare isotope beam by stable S-32. In addition, we describe how the new diagnostic device helped confirm an enhancement of the beam purity prior to beam delivery to experiments.