Your browser doesn't support javascript.
loading
A high risk of postoperative periprosthetic femoral fracture in Dorr type C femurs: a retrospective cohort study with 10-year follow-up data and a preliminary monochromatic image analysis.
Li, Mingyang; Zeng, Yi; Nie, Yong; Liao, Kai; Pei, Fuxing; Yang, Jing; Xie, Huiqi; Shen, Bin.
Afiliação
  • Li M; Department of Orthopedic Surgery and Orthopedic Research Institute, West China Hospital.
  • Zeng Y; Department of Orthopedic Surgery and Orthopedic Research Institute, West China Hospital.
  • Nie Y; Department of Orthopedic Surgery and Orthopedic Research Institute, West China Hospital.
  • Liao K; Department of Radiology, West China Hospital, West China Medical School, Sichuan University, Chengdu, Sichuan Province, People's Republic of China.
  • Pei F; Department of Orthopedic Surgery and Orthopedic Research Institute, West China Hospital.
  • Yang J; Department of Orthopedic Surgery and Orthopedic Research Institute, West China Hospital.
  • Xie H; Department of Orthopedic Surgery and Orthopedic Research Institute, West China Hospital.
  • Shen B; Department of Orthopedic Surgery and Orthopedic Research Institute, West China Hospital.
Int J Surg ; 110(1): 296-305, 2024 Jan 01.
Article em En | MEDLINE | ID: mdl-37830949
ABSTRACT

BACKGROUND:

The authors applied Anatomique Benoist Girard II (ABG II) stems for total hip arthroplasty in some Dorr type C femurs as early attempts. Here, the authors compared the long-term follow-up results between ABG II stems and the 'well-performing' Corail stems and their monochromatic images.

METHODS:

Among 3214 primary total hip arthroplasty records, 43 short ABG II stems and 67 standard-length Corail stems implanted in Dorr type C femurs were eligible and enrolled in this retrospective cohort study, with a mean follow-up of 10.3 years. Revision rates, Harris hip scores, and radiologic signs were compared. Spectral CT scans from a representative sample were obtained, and monochromatic images were reconstructed. A quantitative method was developed to measure the volume of the gap around stems. Patient-specific finite element analysis was conducted to investigate the strains.

RESULTS:

The revision rate of ABG II stems was significantly higher than that of Corail stems (21 vs. 3%, P <0.05). In the monochromatic images, fewer spot-weld signs (2.2 vs. 3.4, P <0.05) and wider gaps around stems (1.64 cm 3 vs. 0.13 cm 3 , P <0.05) were observed on average in the ABG II group. The mean maximum principal strains of the proximal femurs in the ABG II group were close to the yield strains and significantly larger than those in the Corail group (0.0052 vs. 0.0011, P <0.05).

CONCLUSIONS:

There was a high risk of postoperative periprosthetic femoral fracture for ABG II stems in Dorr type C femurs. Monochromatic images provided some insight into the failure mechanism. LEVEL OF EVIDENCE III.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Artroplastia de Quadril / Fraturas Periprotéticas / Fraturas do Fêmur / Prótese de Quadril Limite: Humans Idioma: En Revista: Int J Surg Ano de publicação: 2024 Tipo de documento: Article País de publicação: EEUU / ESTADOS UNIDOS / ESTADOS UNIDOS DA AMERICA / EUA / UNITED STATES / UNITED STATES OF AMERICA / US / USA

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Artroplastia de Quadril / Fraturas Periprotéticas / Fraturas do Fêmur / Prótese de Quadril Limite: Humans Idioma: En Revista: Int J Surg Ano de publicação: 2024 Tipo de documento: Article País de publicação: EEUU / ESTADOS UNIDOS / ESTADOS UNIDOS DA AMERICA / EUA / UNITED STATES / UNITED STATES OF AMERICA / US / USA