Agreement and Reproducibility of Anterior Chamber Angle Measurements between CASIA2 Built-In Software and Human Graders.
Espinoza, Gustavo; Iglesias, Katheriene; Parra, Juan C; Rodriguez-Una, Ignacio; Serrano-Gomez, Sergio; Prada, Angelica M; Galvis, Virgilio.
Afiliação
  • Espinoza G; Centro Oftalmológico Virgilio Galvis, Floridablanca 681004, Santander, Colombia.
  • Iglesias K; Fundación Oftalmológica de Santander, Floridablanca 681004, Santander, Colombia.
  • Parra JC; Facultad de Ciencias de la Salud, Universidad Autónoma de Bucaramanga, Bucaramanga 680002, Santander, Colombia.
  • Rodriguez-Una I; Fundación Oftalmológica de Santander, Floridablanca 681004, Santander, Colombia.
  • Serrano-Gomez S; Fundación Oftalmológica de Santander, Floridablanca 681004, Santander, Colombia.
  • Prada AM; Facultad de Ciencias de la Salud, Universidad Autónoma de Bucaramanga, Bucaramanga 680002, Santander, Colombia.
  • Galvis V; Instituto Universitario Fernández-Vega, Fundación de Investigación Oftalmológica, Universidad de Oviedo, 33012 Oviedo, Spain.
J Clin Med ; 12(19)2023 Oct 06.
Article em En | MEDLINE | ID: mdl-37835024
ABSTRACT

PURPOSE:

This study evaluated the agreement and reproducibility of ACA measurements obtained using the built-in software of the CASIA2 (Version 3G.1) and the measurements derived from expert clinicians.

METHODS:

Healthy volunteers underwent ophthalmological evaluation and AS-OCT examination. ACA measurements derived from automated and manual SS location were obtained using the CASIA2 automated software and clinician identification, respectively. The intraobserver, interobserver reproducibility, CASIA2-human grader reproducibility and CASIA2 repeatability were assessed using intraclass correlation coefficients (ICCs).

RESULTS:

The study examined 58 eyes of 30 participants. The CASIA2 software showed excellent repeatability for all ACA parameters (ICC > 0.84). Intraobserver, interobserver, and CASIA2-human grader reproducibility were also excellent (ICC > 0.87). Interobserver agreement was high, except for nasal TISA500, differing between observers 1 and 2 (p < 0.05). The agreement between CASIA2 measurements and human graders was high, except for nasal TISA500, where observer 1 values were smaller (p < 0.05).

CONCLUSION:

The CASIA2 built-in software reliably measures ACA parameters in healthy individuals, demonstrating high consistency. Although a small difference was observed in nasal TISA500 measurements, interobserver and CASIA2-human grader reproducibility remained excellent. Automated SS detection has the potential to facilitate evaluation and monitoring of primary angle closure disease.