Analysis of Return-to-Zero Error after the First Load of Load Cell.
Sensors (Basel)
; 23(21)2023 Oct 25.
Article
em En
| MEDLINE
| ID: mdl-37960412
ABSTRACT
The return-to-zero error of the resistance strain load cell is most obvious in the first zero-return process during loading and unloading. To improve the accuracy of the load cell, it is necessary to figure out the cause of the error. The influence of the temperature, material, and weld cup were analyzed in this paper. It was concluded that the hysteresis is the main factor affecting the return-to-zero error after the first load. The relationship between hysteresis and zero-return error after first load was obtained by a data fitting algorithm. A method to improve the return-to-zero error after the first load was proposed.
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Idioma:
En
Revista:
Sensors (Basel)
Ano de publicação:
2023
Tipo de documento:
Article
País de afiliação:
China