Biomedical analysis of four fixation systems in treatment of type II traumatic spondylolisthesis of the axis: a finite element analysis.
Med Biol Eng Comput
; 62(3): 843-852, 2024 Mar.
Article
em En
| MEDLINE
| ID: mdl-38052881
This study aimed to compare the properties and safety of self-designed plates in type II traumatic spondylolisthesis of the axis with those of traditional devices via finite element (FE) analysis. We constructed a hangman's fracture FE model from the occipital bone (C0) level to the C3 level. Then, FE models were constructed for the following four fixation systems: an anterior cervical L-shaped plate with four vertebral screws (4-ACLP), or six screws (6-ACLP), an anterior cervical orion plate (ACOP), and a posterior fixation system. A preloaded compressive force of 50 N and a moment of 1.5 N·m were applied to each model under six working conditions. The mobility of the C2/3 segment decreased significantly in four fixation models. In the Mises stress cloud diagram, 4-ACLP showed a better stress distribution in both the bone graft and fixation system than 6-ACLP and ACOP. The resultant force of 4-ACLP was lower but higher than ACOP in axial force. Additionally, the cage in the 4-ACLP configuration experienced the highest stress in the six working conditions. Hence, this novel self-designed plate has the potential to mitigate the operational difficulties, provide sufficient stability, reduce the risk of plate or screw fractures, and improve bone fusion.
Palavras-chave
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Espondilolistese
/
Fraturas da Coluna Vertebral
/
Fraturas Ósseas
Limite:
Humans
Idioma:
En
Revista:
Med Biol Eng Comput
Ano de publicação:
2024
Tipo de documento:
Article
País de afiliação:
China
País de publicação:
Estados Unidos