The Accuracy of Full-Arch Intraoral Optical Impressions (IOS): Clinical Pilot Study of the Influence of the Scan Strategy, Operator, and Intraoral Scanner.
Int J Prosthodont ; 36(6): 689-696, 2023 Dec 18.
Article em En | MEDLINE | ID: mdl-38109389
ABSTRACT

PURPOSE:

To investigate the influence of a novel scanning strategy-using two new intraoral scanner devices with different operators-on the full-arch scanning accuracy for a dentate maxilla. MATERIALS AND

METHODS:

Two scanning strategies, a test and a control strategy, were used to produce full-arch impressions of the dentate maxilla of a study patient. Two intraoral scanning (IOS) devices were used. Five expert operators performed a total of 40 scans. The scan time was recorded for each. A reference model was obtained from the patient's maxillary arch with an analog impression. The model was later scanned with a high-precision laboratory scanner to create a digital reference model (DRM). The scanning accuracy was analyzed with 3D-analysis software using a root mean square (RMS) calculation method, and qualitative analysis was executed using machine learning software.

RESULTS:

The mean RMS result for the test strategy was 82.8 ± 16 µm compared to 81.5 ± 16 µm for the control strategy. The mean RMS results were 84.7 ± 15 µm for Primescan (PS) and 79.6 ± 17 µm for 3Shape (3S). As such, the scanning strategies and IOS devices did not influence the scanning accuracy. Yet, a significant difference was found when the two strategies' scanning times were compared (P = .001), as well as the IOS devices (P = .001). The operator was found to have no influence on the scanning strategy.

CONCLUSIONS:

The accuracy of digital impressions is not influenced by different strategies, devices, or operators, in contrast with the scanning time, which is influenced by both strategies and devices.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Técnica de Moldagem Odontológica / Imageamento Tridimensional Limite: Humans Idioma: En Revista: Int J Prosthodont Assunto da revista: ODONTOLOGIA Ano de publicação: 2023 Tipo de documento: Article País de publicação: Estados Unidos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Técnica de Moldagem Odontológica / Imageamento Tridimensional Limite: Humans Idioma: En Revista: Int J Prosthodont Assunto da revista: ODONTOLOGIA Ano de publicação: 2023 Tipo de documento: Article País de publicação: Estados Unidos