Phantom evaluation of electrical conductivity mapping by MRI: Comparison to vector network analyzer measurements and spatial resolution assessment.
He, Zhongzheng; Lefebvre, Pauline M; Soullié, Paul; Doguet, Martin; Ambarki, Khalid; Chen, Bailiang; Odille, Freddy.
Afiliação
  • He Z; IADI U1254, INSERM and Université de Lorraine, Nancy, France.
  • Lefebvre PM; IADI U1254, INSERM and Université de Lorraine, Nancy, France.
  • Soullié P; IADI U1254, INSERM and Université de Lorraine, Nancy, France.
  • Doguet M; IADI U1254, INSERM and Université de Lorraine, Nancy, France.
  • Ambarki K; BioSerenity, Paris, France.
  • Chen B; Siemens Healthcare SAS, Saint Denis, France.
  • Odille F; IADI U1254, INSERM and Université de Lorraine, Nancy, France.
Magn Reson Med ; 91(6): 2374-2390, 2024 Jun.
Article em En | MEDLINE | ID: mdl-38225861
ABSTRACT

PURPOSE:

To evaluate the performance of various MR electrical properties tomography (MR-EPT) methods at 3 T in terms of absolute quantification and spatial resolution limit for electrical conductivity.

METHODS:

Absolute quantification as well as spatial resolution performance were evaluated on homogeneous phantoms and a phantom with holes of different sizes, respectively. Ground-truth conductivities were measured with an open-ended coaxial probe connected to a vector network analyzer (VNA). Four widely used MR-EPT reconstruction methods were investigated phase-based Helmholtz (PB), phase-based convection-reaction (PB-cr), image-based (IB), and generalized-image-based (GIB). These methods were compared using the same complex images from a 1 mm-isotropic UTE sequence. Alternative transceive phase acquisition sequences were also compared in PB and PB-cr.

RESULTS:

In large homogeneous phantoms, all methods showed a strong correlation with ground truth conductivities (r > 0.99); however, GIB was the best in terms of accuracy, spatial uniformity, and robustness to boundary artifacts. In the resolution phantom, the normalized root-mean-squared error of all methods grew rapidly (>0.40) when the hole size was below 10 mm, with simplified methods (PB and IB), or below 5 mm, with generalized methods (PB-cr and GIB).

CONCLUSION:

VNA measurements are essential to assess the accuracy of MR-EPT. In this study, all tested MR-EPT methods correlated strongly with the VNA measurements. The UTE sequence is recommended for MR-EPT, with the GIB method providing good accuracy for structures down to 5 mm. Structures below 5 mm may still be detected in the conductivity maps, but with significantly lower accuracy.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Processamento de Imagem Assistida por Computador / Encéfalo Idioma: En Revista: Magn Reson Med Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2024 Tipo de documento: Article País de afiliação: França País de publicação: Estados Unidos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Processamento de Imagem Assistida por Computador / Encéfalo Idioma: En Revista: Magn Reson Med Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2024 Tipo de documento: Article País de afiliação: França País de publicação: Estados Unidos