Correction of axial length measurement error by IOLMaster 700 could improve refractive prediction accuracy in silicone oil-filled eyes.
Acta Ophthalmol
; 102(5): e718-e726, 2024 Aug.
Article
em En
| MEDLINE
| ID: mdl-38235601
ABSTRACT
PURPOSE:
To determine whether correcting the axial length (AL) measurement error of the IOLMaster 700 could improve the refractive prediction accuracy in silicone oil-filled eyes.METHODS:
This study included 265 cataract patients (265 eyes) with silicone oil tamponade who were scheduled for phacoemulsification with intraocular lens (IOL) implantation. The performances of various formulas, including Barrett Universal II, Emmetropia Verifying Optical, Hoffer-QST, Kane, Ladas Super Formula, Pearl-DGS, Radial Basis Function and traditional formulas (Haigis, Hoffer Q, Holladay 1 and SRK/T), were evaluated. The refractive prediction errors (PE) calculated with measured AL (ALmeas) and corrected AL with silicone oil adjustment (SOAL) were compared. Subgroup analysis was performed based on the ALmeas (<23 mm; 23-26 mm; ≥26 mm).RESULTS:
Using SOAL significantly reduced the hyperopic PE of formulas when compared to ALmeas (-0.05 to 0.17 D vs 0.15 to 0.38 D, p < 0.001). After applying AL correction, all formulas showed a lower mean absolute PE (0.47-0.57 D vs 0.50-0.69 D). The percentage of eyes within ±1.0 D of PE increased from 84.91%-88.68% to 89.81%-91.32% for new formulas and from 78.11%-83.40% to 85.66%-88.68% for traditional formulas, with the use of SOAL. Subgroup analysis showed that the majority of formulas with SOAL in prediction accuracy for eyes with an AL ≥26 mm (p < 0.05).CONCLUSIONS:
The refractive prediction accuracy in silicone oil-filled eyes was improved by correcting the AL measurement error of the IOLMaster 700, especially for long eyes.Palavras-chave
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Refração Ocular
/
Óleos de Silicone
/
Facoemulsificação
/
Comprimento Axial do Olho
Tipo de estudo:
Prognostic_studies
/
Risk_factors_studies
Limite:
Aged
/
Aged80
/
Female
/
Humans
/
Male
/
Middle aged
Idioma:
En
Revista:
Acta Ophthalmol
Assunto da revista:
OFTALMOLOGIA
Ano de publicação:
2024
Tipo de documento:
Article
País de afiliação:
China
País de publicação:
Reino Unido