Confounders of Ultrasound Attenuation Imaging in a Linear Probe Using the Canon Aplio i800 System: A Phantom Study.
Diagnostics (Basel)
; 14(3)2024 Jan 26.
Article
em En
| MEDLINE
| ID: mdl-38337786
ABSTRACT
There have been studies showing attenuation imaging (ATI) with ultrasound as an approach to diagnose liver diseases such as steatosis or cirrhosis. So far, this technique has only been used on a convex probe. The goal of the study was to investigate the feasibility of ATI measurements using the linear array on a canon Aplio i800 scanner on certified phantoms. Three certified liver tissue attenuation phantoms were measured in five different positions using a linear probe. The effects of positioning and depth were explored and compared. The values were compared to the certified expected value for each phantom as well as the different measurement values for each measurement position. The ATI measurements on phantoms showed significant effect for the different probe positions and region of interest (ROI) depths. Values taken in the center with the probe perpendicular to the phantom were closest to certified values. Median values at 2.5-4.5 cm depth for phantoms 1 and 2 and 0.5-2.5 cm for phantom 3 were comparable with certified values. Measurements taken at a depth greater than 6 cm in any position were the least representative of the certified values (p-value < 0.01) and had the widest range throughout the different sessions. ATI measurements can be performed with the linear probe in phantoms; however, careful consideration should be given to depth dependency, as it can significantly affect measurement values. Remaining measurements at various depths within the 0.5-6.0 cm range showed deviation from the certified values of approximately 25%.
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Idioma:
En
Revista:
Diagnostics (Basel)
Ano de publicação:
2024
Tipo de documento:
Article
País de afiliação:
Suíça
País de publicação:
Suíça