Beyond helium: hydrogen as a carrier gas in multiresidue pesticide analysis in fruits and vegetables by GC-MS/MS.
Cutillas, Víctor; García-Gallego, Guillermo; Murcia-Morales, María; Ferrer, Carmen; Fernández-Alba, Amadeo R.
Afiliação
  • Cutillas V; European Union Reference Laboratory for Pesticide Residues in Fruit & Vegetables, Agrifood Campus of International Excellence (ceiA3), University of Almeria, Ctra. Sacramento S/N, La Cañada de San Urbano, 04120, Almería, Spain. amadeo@ual.es.
  • García-Gallego G; European Union Reference Laboratory for Pesticide Residues in Fruit & Vegetables, Agrifood Campus of International Excellence (ceiA3), University of Almeria, Ctra. Sacramento S/N, La Cañada de San Urbano, 04120, Almería, Spain. amadeo@ual.es.
  • Murcia-Morales M; European Union Reference Laboratory for Pesticide Residues in Fruit & Vegetables, Agrifood Campus of International Excellence (ceiA3), University of Almeria, Ctra. Sacramento S/N, La Cañada de San Urbano, 04120, Almería, Spain. amadeo@ual.es.
  • Ferrer C; European Union Reference Laboratory for Pesticide Residues in Fruit & Vegetables, Agrifood Campus of International Excellence (ceiA3), University of Almeria, Ctra. Sacramento S/N, La Cañada de San Urbano, 04120, Almería, Spain. amadeo@ual.es.
  • Fernández-Alba AR; European Union Reference Laboratory for Pesticide Residues in Fruit & Vegetables, Agrifood Campus of International Excellence (ceiA3), University of Almeria, Ctra. Sacramento S/N, La Cañada de San Urbano, 04120, Almería, Spain. amadeo@ual.es.
Anal Methods ; 16(11): 1564-1569, 2024 03 14.
Article em En | MEDLINE | ID: mdl-38406836
ABSTRACT
In this comprehensive study, we evaluated the feasibility of using hydrogen instead of helium as a carrier gas in a GC-MS/MS system for pesticide residue analysis, spanning three matrices pepper, tomato, and zucchini. Initial assessments focused on the ion source's chemical inertness, employing nitrobenzene as a benchmark to monitor the hydrogenation process. A method with a duration of less than 12 minutes was developed, achieving good chromatographic peak resolution attributable to the enhanced chromatographic performance of hydrogen as a carrier gas. The study emphasized the optimization of system parameters, testing various ion source temperatures, detector voltages, and injection volumes. Sensitivity assessments, based on the DG-SANTE criteria, indicated that the majority of compounds were identifiable at a concentration of 5 µg kg-1 (81% in tomato, 84% in pepper and 73% in zucchini). Detailed validation for reproducibility, matrix effects, and linearity across 150 pesticides unveiled generally favorable outcomes, with a notable majority of compounds displaying low matrix effects, satisfactory linearity ranges and good reproducibility with most compounds returning a relative standard deviation (RSD) below 10%. When applied to 15 real samples, the hydrogen-based system's performance was juxtaposed against a helium-based counterpart, revealing that results are very comparable between both systems. This comparative approach highlights hydrogen's potential as a reliable and efficient carrier gas in pesticide residue analysis for routine food control laboratories, overcoming difficulties resulting from the lack of helium supplies.