Your browser doesn't support javascript.
loading
A Novel Approach for Position Verification and Dose Calculation through Local MVCT Reconstruction.
Zhang, Jun; Chen, Zerui; Lei, Yuxin; Wen, Junhai.
Afiliação
  • Zhang J; College of Computer Science and Technology, Taiyuan University of Technology, Taiyuan 030024, China.
  • Chen Z; School of Life Science, Beijing Institute of Technology, Beijing 100081, China.
  • Lei Y; School of Life Science, Beijing Institute of Technology, Beijing 100081, China.
  • Wen J; School of Life Science, Beijing Institute of Technology, Beijing 100081, China.
Diagnostics (Basel) ; 14(5)2024 Feb 23.
Article em En | MEDLINE | ID: mdl-38472954
ABSTRACT
Traditional positioning verification using cone-beam computed tomography (CBCT) may incur errors due to potential misalignments between the isocenter of CBCT and the treatment beams in radiotherapy. This study introduces an innovative method for verifying patient positioning in radiotherapy. Initially, the transmission images from an electronic portal imaging device (EPID) are acquired from 10 distinct angles. Utilizing the ART-TV algorithm, a sparse reconstruction of local megavoltage computed tomography (MVCT) is performed. Subsequently, this MVCT is aligned with the planning CT via a three-dimensional mutual information registration technique, pinpointing any patient-positioning discrepancies and facilitating corrective adjustments to the treatment setup. Notably, this approach employs the same radiation source as used in treatment to obtain three-dimensional images, thereby circumventing errors stemming from misalignment between the isocenter of CBCT and the accelerator. The registration process requires only 10 EPID images, and the dose absorbed during this process is included in the total dose calculation. The results show that our method's reconstructed MVCT images fulfill the requirements for registration, and the registration algorithm accurately detects positioning errors, thus allowing for adjustments in the patient's treatment position and precise calculation of the absorbed dose.
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Revista: Diagnostics (Basel) Ano de publicação: 2024 Tipo de documento: Article País de afiliação: China

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Revista: Diagnostics (Basel) Ano de publicação: 2024 Tipo de documento: Article País de afiliação: China