Comparing region of interest selection and whole-field analysis for measurement of ciliary beat frequency in high-speed video analysis.
Abo, Miki; Imamura, Kousuke; Hosogi, Shigekuni; Kobayashi, Takafumi; Takeda, Yoshihiro; Kase, Kazumasa; Koba, Hayato; Watanabe, Satoshi; Ohkura, Noriyuki; Hara, Johsuke; Yano, Seiji.
Afiliação
  • Abo M; Kanazawa University Health Service Center, Kakuma-machi, Kanazawa, Ishikawa, 920-1192, Japan; Department of Respiratory Medicine, Kanazawa University Hospital, 13-1 Takara-machi, Kanazawa, Ishikawa, 920-8641, Japan. Electronic address: abomiki@staff.kanazwa-u.ac.jp.
  • Imamura K; Department of Electrical, Information and Communication Engineering, Kanazawa University, Kakuma-machi, Kanazawa, Ishikawa, 920-1192, Japan.
  • Hosogi S; Department of Clinical and Translational Physiology, Kyoto Pharmaceutical University, Kyoto, 607-8414, Japan.
  • Kobayashi T; Division of Medical Oncology, Cancer Research Institute, Kanazawa University, Kanazawa, Ishikawa, 920-0934, Japan.
  • Takeda Y; Department of Respiratory Medicine, Kanazawa University Hospital, 13-1 Takara-machi, Kanazawa, Ishikawa, 920-8641, Japan.
  • Kase K; Department of Respiratory Medicine, Kanazawa University Hospital, 13-1 Takara-machi, Kanazawa, Ishikawa, 920-8641, Japan.
  • Koba H; Department of Respiratory Medicine, Kanazawa University Hospital, 13-1 Takara-machi, Kanazawa, Ishikawa, 920-8641, Japan.
  • Watanabe S; Department of Respiratory Medicine, Kanazawa University Hospital, 13-1 Takara-machi, Kanazawa, Ishikawa, 920-8641, Japan.
  • Ohkura N; Department of Respiratory Medicine, Kanazawa University Hospital, 13-1 Takara-machi, Kanazawa, Ishikawa, 920-8641, Japan.
  • Hara J; Department of Respiratory Medicine, Kanazawa University Hospital, 13-1 Takara-machi, Kanazawa, Ishikawa, 920-8641, Japan.
  • Yano S; Department of Respiratory Medicine, Kanazawa University Hospital, 13-1 Takara-machi, Kanazawa, Ishikawa, 920-8641, Japan; Division of Medical Oncology, Cancer Research Institute, Kanazawa University, Kanazawa, Ishikawa, 920-0934, Japan; Nano Life Science Institute (WPI-NanoLSI), Kanazawa University,
Respir Investig ; 62(3): 419-425, 2024 May.
Article em En | MEDLINE | ID: mdl-38489921
ABSTRACT

BACKGROUND:

Ciliary beat frequency (CBF) is crucial in mucociliary clearance. High-speed video analysis (HSVA) is commonly used to measure CBF but lacks standardization. We compared visual observation and computer-assisted calculation using fast Fourier transformation (FFT) in freshly collected bronchial ciliary epithelial cells and cultured cells.

METHODS:

Bronchial epithelial cells were obtained from 12 patients who required bronchoscopic examination. Eighty-five videos of ciliary movement of freshly collected and cultured cells were recorded and used to calculate CBF using manual observation, region of interest (ROI) selection, and whole-field analysis.

RESULTS:

CBF measured by the ROI selection method strongly correlated with that measured using manual observation, especially in freshly collected cells. However, 27.8% of the manual observation method values were doubled in the ROI selection method, probably because a round trip of cilia was calculated as two cycles and needed to be corrected to 1/2 value. Upon increasing the number of ROIs, the results of the ROI selection method came closer to that of WFA.

CONCLUSIONS:

Computer-assisted calculation using FFT can aid in measuring CBF; however, current methods require visual confirmation. Further automated evaluation techniques are needed to establish more standardized and generalized CBF measurement methods using HSVA.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Brônquios / Depuração Mucociliar Limite: Humans Idioma: En Revista: Respir Investig Ano de publicação: 2024 Tipo de documento: Article País de publicação: HOLANDA / HOLLAND / NETHERLANDS / NL / PAISES BAJOS / THE NETHERLANDS

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Brônquios / Depuração Mucociliar Limite: Humans Idioma: En Revista: Respir Investig Ano de publicação: 2024 Tipo de documento: Article País de publicação: HOLANDA / HOLLAND / NETHERLANDS / NL / PAISES BAJOS / THE NETHERLANDS