Feasibility study of radioactivity estimation of 99mTc and 123I-labeled radiopharmaceuticals using shielded syringes.
Radiol Phys Technol
; 17(2): 396-401, 2024 Jun.
Article
em En
| MEDLINE
| ID: mdl-38517669
ABSTRACT
This study investigates the feasibility of estimating the radioactivity of radiopharmaceuticals using shielded syringes. The radioactivities of 99mTc-MDP, 99mTc-HMDP, 99mTc-ECD, 99mTc-MAG3, and 123I-IMP were measured using a dose calibrator. Correlation coefficients and regression equations were obtained from the radioactivity in the shielded and unshielded syringes. 99mTc-MDP was also measured for residual radioactivity after the administration. The correlation coefficients of 99mTc-MDP, 99mTc-HMDP, 99mTc-ECD, 99mTc-MAG3, and 123I-IMP were rs = 0.9998, rs = 0.9997, rs = 0.9999, rs = 0.9998, and rs = 0.9888, respectively. The regression equations were y = 0.0364x + 0.0913, y = 0.0349x + 0.0273, y = 0.0343x - 0.0018, y = 0.0522x + 0.1215, and y = 0.0383x + 0.0058, respectively. The correlation coefficient for the residual radioactivity of 99mTc-MDP was rs = 0.9887 and the regression equation was y = 0.1505x + 0.0853. The radioactivity of 99mTc- and 123I-labeled radiopharmaceuticals in shielded syringes was accurately measured. It was suggested that the measuring shielded syringes could provide an estimate of the actual radioactivity.
Palavras-chave
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Seringas
/
Estudos de Viabilidade
/
Tecnécio
/
Compostos Radiofarmacêuticos
/
Radioisótopos do Iodo
Idioma:
En
Revista:
Radiol Phys Technol
Assunto da revista:
BIOFISICA
/
RADIOLOGIA
Ano de publicação:
2024
Tipo de documento:
Article
País de afiliação:
Japão
País de publicação:
Japão