Your browser doesn't support javascript.
loading
Differences Between Simulated Keratometry and Total Corneal Power in Eyes With Keratoconus and a Formula to Improve IOL Power Calculation Results.
J Refract Surg ; 40(4): e253-e259, 2024 Apr.
Article em En | MEDLINE | ID: mdl-38593262
ABSTRACT

PURPOSE:

To compare simulated keratometry (SimK) and total corneal power (TCP) in keratoconic eyes, to determine whether the differences are systematic and predictable and to evaluate an adjusted TCP-based formula for intraocular lens (IOL) power calculation.

METHODS:

In a consecutive series of keratoconic eyes, measurements of SimK, TCP, posterior keratometry, and anterior and posterior corneal asphericities (Q-values) were retrospectively collected. The difference between SimK and TCP was linearly correlated to the biometric parameters. In a separate sample of keratoconic eyes that had undergone cataract surgery, IOL power was calculated with the Barrett Universal II, Hoffer QST, Holladay 1, Kane, and SRK/T formulas using the SimK and an adjusted TCP power. The respective prediction errors were calculated.

RESULTS:

A total of 382 keratoconic eyes (271 patients) were enrolled. An increasing overestimation of SimK by TCP was detected from stage I to III, with a significant correlation between the SimK and TCP difference and SimK in the whole sample (P < .0001, r2 = 0.1322). Approximately 7% of cases presented an underestimation of SimK by TCP. IOL power calculation with the adjusted TCP improved outcomes, achieving a maximum of 80% of eyes with a prediction error within ±0.50 diopters with the Hoffer QST, Holladay 1, and Kane formulas.

CONCLUSIONS:

Overall, SimK overestimated TCP. Such a difference could not be predicted by any variable. The proposed TCP-adjustment formula (TCPadj = TCP + 0.56 diopters) in keratoconic eyes for IOL power calculation might be valuable for improving refractive outcomes. [J Refract Surg. 2024;40(4)e253-e259.].
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Facoemulsificação / Ceratocone / Lentes Intraoculares Limite: Humans Idioma: En Revista: J Refract Surg Assunto da revista: OFTALMOLOGIA Ano de publicação: 2024 Tipo de documento: Article País de publicação: Estados Unidos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Facoemulsificação / Ceratocone / Lentes Intraoculares Limite: Humans Idioma: En Revista: J Refract Surg Assunto da revista: OFTALMOLOGIA Ano de publicação: 2024 Tipo de documento: Article País de publicação: Estados Unidos