CoaguChek and Coag-Sense PT2 Meter Point of Care INR Device Validation.
McBane, Robert D; O'Connor, Carlee; Lutz, Jennifer; Blanco, Jennifer; Hartman, Levi A; Kramer, Anna; Uy, Josephine; Schumann, Traci; Hartung, Katherine; Luker, Michael; Hodge, David; Santrach, Paula; Karon, Brad S.
Afiliação
  • McBane RD; Gonda Vascular Center, Mayo Clinic, Rochester, MN, USA; Cardiovascular Department, Mayo Clinic, Rochester, MN, USA. Electronic address: mcbane.robert@mayo.edu.
  • O'Connor C; Anticoagulation Services, Mayo Clinic, Eau Claire, WI, USA.
  • Lutz J; Anticoagulation Services, Mayo Clinic, Scottsdale, AZ, USA.
  • Blanco J; Gonda Vascular Center, Mayo Clinic, Rochester, MN, USA.
  • Hartman LA; Department of Laboratory Medicine, Mayo Clinic, Rochester, MN, USA.
  • Kramer A; Quality Management Services, Mayo Clinic, La Crosse, WI, USA.
  • Uy J; Department of Laboratory Medicine, Mayo Clinic, Scottsdale, AZ, USA.
  • Schumann T; Department of Laboratory Medicine, Mayo Clinic, Rochester, MN, USA.
  • Hartung K; Department of Laboratory Medicine, Mayo Clinic, Rochester, MN, USA.
  • Luker M; Department on Internal Medicine, Mayo Clinic, Rochester, MN, USA.
  • Hodge D; Department of Quantitative Health Sciences, Mayo Clinic, Jacksonville, FL, USA.
  • Santrach P; Department of Laboratory Medicine, Mayo Clinic, Rochester, MN, USA.
  • Karon BS; Department of Laboratory Medicine, Mayo Clinic, Rochester, MN, USA.
Mayo Clin Proc ; 99(7): 1091-1100, 2024 Jul.
Article em En | MEDLINE | ID: mdl-38661594
ABSTRACT

OBJECTIVE:

To standardize international normalized ratio (INR) measurements and improve data integrity by enabling electronic result transmission for warfarin monitoring, two point-of-care (POC) devices were evaluated against an internal plasma INR reference method.

METHODS:

A multicenter study was pursued (January 24, 2022, through October 19, 2022) to compare concordance of two commercially available POC devices, Coag-Sense PT2 Meter (Coag-Sense) and CoaguChek XS Pro and Plus devices (CoaguChek), against an internal plasma INR method among patients treated with warfarin. Bias and linear regression analysis were assessed for these devices including dosing decision accuracy compared with plasma INR reference.

RESULTS:

Two hundred ninety-nine patients treated with warfarin across three Mayo Clinic sites agreed to participate. Atrial fibrillation (n=191, 63.9%), venous thromboembolism (n=65; 21.7%), and heart valve prosthesis (n=46; 15.4%) were common anticoagulant indications with a 2.5 INR target for 280 (93.6%) of patients. For the CoaguChek devices, 243 (81.3%) of values fell within 0.2 INR units with plasma INR referent and 285 (95.3%) within 0.4 units (R2=0.93). For the Coag-Sense device, 102 (34.1%) of values fell within 0.2 INR units and 180 (60.2%) within 0.4 INR units of plasma INR values, (R2=0.83; P<.0001). Using the plasma INR as the gold standard, appropriate dosing recommendations would have occurred for 292 (97.7%) of the CoaguChek and 244 (81.6%) of the Coag-Sense results.

CONCLUSION:

Compared with a plasma referent, INR values obtained from the CoaguChek devices exhibited less systematic bias compared with Coag-Sense measures. This translates to a greater percentage of concordant management decisions between POC and laboratory INR methods.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Varfarina / Monitoramento de Medicamentos / Sistemas Automatizados de Assistência Junto ao Leito / Coeficiente Internacional Normatizado / Anticoagulantes Limite: Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Mayo Clin Proc Ano de publicação: 2024 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Varfarina / Monitoramento de Medicamentos / Sistemas Automatizados de Assistência Junto ao Leito / Coeficiente Internacional Normatizado / Anticoagulantes Limite: Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Mayo Clin Proc Ano de publicação: 2024 Tipo de documento: Article