Your browser doesn't support javascript.
loading
Multicenter comparison using two AI stroke CT perfusion software packages for determining thrombectomy eligibility.
Alwood, Benjamin T; Meyer, Dawn M; Ionita, Chip; Snyder, Kenneth V; Santos, Roberta; Perrotta, Lindsey; Crooks, Ryan; Van Orden, Kimberlee; Torres, Dolores; Poynor, Briana; Pham, Nhan; Kelly, Sophie; Meyer, Brett C; Bolar, Divya S.
Afiliação
  • Alwood BT; Department of Vascular Neurology, University of Florida, Jacksonville, FL, United States; University of California San Diego Stroke Center, University of California San Diego, San Diego, CA, United States. Electronic address: benjamin.alwood@jax.ufl.edu.
  • Meyer DM; University of California San Diego Stroke Center, University of California San Diego, San Diego, CA, United States.
  • Ionita C; Department of Biomedical Engineering and Neurosurgery, University at Buffalo, Buffalo NY, United States.
  • Snyder KV; Department of Biomedical Engineering and Neurosurgery, University at Buffalo, Buffalo NY, United States.
  • Santos R; Department of Vascular Neurology, University of Florida, Jacksonville, FL, United States.
  • Perrotta L; Department of Vascular Neurology, University of Florida, Jacksonville, FL, United States.
  • Crooks R; Department of Vascular Neurology, University of Florida, Jacksonville, FL, United States.
  • Van Orden K; University of California San Diego Stroke Center, University of California San Diego, San Diego, CA, United States.
  • Torres D; University of California San Diego Stroke Center, University of California San Diego, San Diego, CA, United States.
  • Poynor B; University of California San Diego Stroke Center, University of California San Diego, San Diego, CA, United States.
  • Pham N; Department of Radiology, University of California San Diego, San Diego, CA, United States.
  • Kelly S; Department of Radiology, University of California San Diego, San Diego, CA, United States.
  • Meyer BC; University of California San Diego Stroke Center, University of California San Diego, San Diego, CA, United States.
  • Bolar DS; Department of Radiology, University of California San Diego, San Diego, CA, United States; Center for Functional MRI, University of California San Diego, San Diego, CA, United States.
J Stroke Cerebrovasc Dis ; 33(7): 107750, 2024 Jul.
Article em En | MEDLINE | ID: mdl-38703875
ABSTRACT

BACKGROUND:

Stroke AI platforms assess infarcted core and potentially salvageable tissue (penumbra) to identify patients suitable for mechanical thrombectomy. Few studies have compared outputs of these platforms, and none have been multicenter or considered NIHSS or scanner/protocol differences. Our objective was to compare volume estimates and thrombectomy eligibility from two widely used CT perfusion (CTP) packages, Viz.ai and RAPID.AI, in a large multicenter cohort.

METHODS:

We analyzed CTP data of acute stroke patients with large vessel occlusion (LVO) from four institutions. Core and penumbra volumes were estimated by each software and DEFUSE-3 thrombectomy eligibility assessed. Results between software packages were compared and categorized by NIHSS score, scanner manufacturer/model, and institution.

RESULTS:

Primary analysis of 362 cases found statistically significant differences in both software's volume estimations, with subgroup analysis showing these differences were driven by results from a single scanner model, the Canon Aquilion One. Viz.ai provided larger estimates with mean differences of 8cc and 18cc for core and penumbra, respectively (p<0.001). NIHSS subgroup analysis also showed systematically larger Viz.ai volumes (p<0.001). Despite volume differences, a significant difference in thrombectomy eligibility was not found. Additional subgroup analysis showed significant differences in penumbra volume for the Phillips Ingenuity scanner, and thrombectomy eligibility for the Canon Aquilion One scanner at one center (7 % increased eligibility with Viz.ai, p=0.03).

CONCLUSIONS:

Despite systematic differences in core and penumbra volume estimates between Viz.ai and RAPID.AI, DEFUSE-3 eligibility was not statistically different in primary or NIHSS subgroup analysis. A DEFUSE-3 eligibility difference, however, was seen on one scanner at one institution, suggesting scanner model and local CTP protocols can influence performance and cause discrepancies in thrombectomy eligibility. We thus recommend centers discuss optimal scanning protocols with software vendors and scanner manufacturers to maximize CTP accuracy.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Software / Interpretação de Imagem Radiográfica Assistida por Computador / Circulação Cerebrovascular / Valor Preditivo dos Testes / Trombectomia / Seleção de Pacientes / Imagem de Perfusão Limite: Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: J Stroke Cerebrovasc Dis Assunto da revista: ANGIOLOGIA / CEREBRO Ano de publicação: 2024 Tipo de documento: Article País de publicação: Estados Unidos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Software / Interpretação de Imagem Radiográfica Assistida por Computador / Circulação Cerebrovascular / Valor Preditivo dos Testes / Trombectomia / Seleção de Pacientes / Imagem de Perfusão Limite: Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: J Stroke Cerebrovasc Dis Assunto da revista: ANGIOLOGIA / CEREBRO Ano de publicação: 2024 Tipo de documento: Article País de publicação: Estados Unidos