Evolution of a laboratory mechanomyograph.
Wedemeyer, Zain; Bowdle, Andrew; Jelacic, Srdjan; Lopez, Aidan; Silliman, Willis; Michaelsen, Kelly E.
Afiliação
  • Wedemeyer Z; Department of Anesthesiology, University of Washington, Box 356540, 1959 NE Pacific Street, Seattle, WA, 98195-6540, USA.
  • Bowdle A; Department of Anesthesiology, University of Washington, Box 356540, 1959 NE Pacific Street, Seattle, WA, 98195-6540, USA.
  • Jelacic S; Department of Anesthesiology, University of Washington, Box 356540, 1959 NE Pacific Street, Seattle, WA, 98195-6540, USA.
  • Lopez A; Department of Anesthesiology, University of Washington, Box 356540, 1959 NE Pacific Street, Seattle, WA, 98195-6540, USA.
  • Silliman W; Department of Anesthesiology, University of Washington, Box 356540, 1959 NE Pacific Street, Seattle, WA, 98195-6540, USA.
  • Michaelsen KE; Department of Anesthesiology, University of Washington, Box 356540, 1959 NE Pacific Street, Seattle, WA, 98195-6540, USA. kellyem@uw.edu.
J Clin Monit Comput ; 2024 May 17.
Article em En | MEDLINE | ID: mdl-38758405
ABSTRACT
Mechanomyography is currently the accepted laboratory reference standard for quantitative neuromuscular blockade monitoring. Mechanomyographs are not commercially available. Previously, a mechanomyograph was built by our laboratory and used in several clinical studies. It was subsequently redesigned to improve its usability and functionality and to accommodate a wider range of hand sizes and shapes using an iterative design process. Each version of the redesigned device was initially tested for usability and functionality in the lab with the investigators as subjects without electrical stimulation. The redesigned devices were then assessed on patients undergoing elective surgery under general anesthesia without neuromuscular blocking drugs. Since the patients were not paralyzed, the expected train-of-four ratio was 1.0. The device accuracy and precision were represented by the train-of-four ratio mean and standard deviation. If issues with the device's useability or functionality were discovered, changes were made, and the redesign processes repeated. The final mechanomyograph design was used to collect 2,362 train-of-four ratios from 21 patients. The mean and standard deviation of the train-of-four ratios were 0.99 ± 0.030. Additionally, the final mechanomyograph design was easier to use and adjust than the original design and fit a wider range of hand sizes. The final design also reduced the frequency of adjustments and the time needed for adjustments, facilitating data collection during a surgical procedure.
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Revista: J Clin Monit Comput Assunto da revista: INFORMATICA MEDICA / MEDICINA Ano de publicação: 2024 Tipo de documento: Article País de afiliação: Estados Unidos País de publicação: Holanda