Toward Next-Generation Molecular Imaging with a Clinical Low-Field (0.064 T) Point-of-Care MRI Scanner.
Anal Chem
; 96(25): 10348-10355, 2024 06 25.
Article
em En
| MEDLINE
| ID: mdl-38857182
ABSTRACT
Low-field (LF) MRI promises soft-tissue imaging without the expensive, immobile magnets of clinical scanners but generally suffers from limited detection sensitivity and contrast. The sensitivity boost provided by hyperpolarization can thus be highly synergistic with LF MRI. Initial efforts to integrate a continuous-bubbling SABRE (signal amplification by reversible exchange) hyperpolarization setup with a portable, point-of-care 64 mT clinical MRI scanner are reported. Results from 1H SABRE MRI of pyrazine and nicotinamide are compared with those of benchtop NMR spectroscopy. Comparison with MRI signals from samples with known H2O/D2O ratios allowed quantification of the SABRE enhancements of imaged samples with various substrate concentrations (down to 3 mM). Respective limits of detection and quantification of 3.3 and 10.1 mM were determined with pyrazine 1H polarization (PH) enhancements of â¼1900 (PH â¼0.04%), supporting ongoing and envisioned efforts to realize SABRE-enabled MRI-based molecular imaging.
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Pirazinas
/
Imageamento por Ressonância Magnética
/
Niacinamida
/
Sistemas Automatizados de Assistência Junto ao Leito
/
Imagem Molecular
Limite:
Humans
Idioma:
En
Revista:
Anal Chem
Ano de publicação:
2024
Tipo de documento:
Article
País de afiliação:
Estados Unidos
País de publicação:
Estados Unidos