ABSTRACT
In photoacoustic tomography (PAT), acoustic inversion aims to recover the spatial distribution of light energy deposition within the imaging object from the signals captured by detectors. To achieve quantitative imaging, optical inversion is further employed to derive absorption coefficient (AC) images. However, limitations such as restricted detection angles and inherent noise lead to substantial artifacts and degradation in the quality of PAT images, consequently affecting the accuracy of optical inversion results. In this study, we propose a directional total variation constrained optical inversion model to reconstruct the AC image. By incorporating anatomy prior information into the optical inversion process, our method can effectively suppress artifacts in AC images while maintaining structural integrity. Simulation, phantom, and in vivo experimental results demonstrate that our method significantly improves the reconstructed AC image quality. Our method provides a reliable foundation for achieving high-quality quantitative PAT imaging.