Research on Detection Efficiency of Imaging Plates for Alpha Particles Using Two Types of Imaging Plate.
Health Phys
; 127(4): 481-489, 2024 Oct 01.
Article
em En
| MEDLINE
| ID: mdl-38888445
ABSTRACT
ABSTRACT Imaging plates can measure isotopes with alpha decay (such as radon and its progeny, americium, and so on). However, the detection efficiency of imaging plates is affected by alpha particle energy, types of imaging plates, and the overlapping effect. In this study, simulations were performed to analyze the relationship between detection efficiency and these three influence factors. The research findings suggest that BAS-TR and BAS-MS are well-suited for the detection of alpha particles with energy levels below 6.83 MeV and above, respectively. The track overlap effect correction method proposed in this study is applicable to both BAS-TR and BAS-MS image plates. The measurement results of radon progeny demonstrate that the correction method enhances the detection efficiency from 0.203 to 0.288. This study presents a valuable approach for selecting the appropriate image plate and correcting the track overlap effect in the measurement of alpha radioactive material concentration and other related information.