Your browser doesn't support javascript.
loading
Total body irradiation: A transition from a Co-60 treatment unit to an IMRT lateral-field extended-SAD technique.
Kolokotronis, Anastasia; Brunet-Benkhoucha, Malik; Roussin, Étienne; St-Pierre, Julie; Marchand, Eve-Lyne; Bernard, Maryse.
Afiliação
  • Kolokotronis A; Département de Radio-oncologie, CIUSSS de l'Est-de-L'Île-de-Montréal, Hôpital Maisonneuve-Rosemont, Montréal, Quebec, Canada.
  • Brunet-Benkhoucha M; Département de Radio-oncologie, CIUSSS de l'Est-de-L'Île-de-Montréal, Hôpital Maisonneuve-Rosemont, Montréal, Quebec, Canada.
  • Roussin É; Département de Radio-oncologie, CIUSSS de l'Est-de-L'Île-de-Montréal, Hôpital Maisonneuve-Rosemont, Montréal, Quebec, Canada.
  • St-Pierre J; Département de Radio-oncologie, CIUSSS de l'Est-de-L'Île-de-Montréal, Hôpital Maisonneuve-Rosemont, Montréal, Quebec, Canada.
  • Marchand EL; Département de Radio-oncologie, CIUSSS de l'Est-de-L'Île-de-Montréal, Hôpital Maisonneuve-Rosemont, Montréal, Quebec, Canada.
  • Bernard M; Département de Radio-oncologie, CIUSSS de l'Est-de-L'Île-de-Montréal, Hôpital Maisonneuve-Rosemont, Montréal, Quebec, Canada.
J Appl Clin Med Phys ; 25(9): e14430, 2024 Sep.
Article em En | MEDLINE | ID: mdl-38952071
ABSTRACT

PURPOSE:

The purpose of this work was to detail our center's experience in transitioning from a Co-60 treatment technique to an intensity modulated radiation therapy (IMRT) based lateral-field extended source-to-axis distance (e-SAD) technique for total body irradiation (TBI). MATERIALS AND

METHODS:

An existing beam model in RayStation v.10A was validated for the use of e-SAD TBI treatments. Data were acquired with an Elekta Synergy linear accelerator (LINAC) at an extended source-to-surface distance of 365 cm with an 18 MV beam. Beam model validation measurements included percentage depth dose (PDD), profile data, surface dose, build-up region and transmission measurements. End-to-end testing was carried out using an anthropomorphic phantom. Treatments were performed in a supine position in a whole-body Vac-Lok at an e-SAD of 400 cm with a beam spoiler 10 cm from the couch. Planning was achieved using IMRT, where multi-leaf collimators were used to modulate the beam and shield the organs at risk. Beam's eye view projection images were used for in-room patient positioning and in-vivo dosimetry was performed for every treatment.

RESULTS:

The percent difference between the measured and calculated PDD and profiles was less than 2% at all locations. Surface dose was 83.8% of the maximum dose with the beam spoiler at a 10 cm distance from the phantom. The largest percent difference between the treatment planning system (TPS) and measured data within the anthropomorphic phantom was approximately 2%. In-vivo dosimetry measurements yielded results within the 5% institutional threshold.

CONCLUSION:

In 2022, 17 patients were successfully treated using the new IMRT-based lateral-field e-SAD TBI technique. The resulting clinical plans respected the institutional standard. The commissioning process, as well as the treatment planning and delivery aspects were described in this work with the intention of supporting other clinics in implementing this treatment method.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Aceleradores de Partículas / Dosagem Radioterapêutica / Planejamento da Radioterapia Assistida por Computador / Irradiação Corporal Total / Radioisótopos de Cobalto / Imagens de Fantasmas / Radioterapia de Intensidade Modulada / Órgãos em Risco Limite: Humans Idioma: En Revista: J Appl Clin Med Phys Assunto da revista: BIOFISICA Ano de publicação: 2024 Tipo de documento: Article País de afiliação: Canadá País de publicação: Estados Unidos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Aceleradores de Partículas / Dosagem Radioterapêutica / Planejamento da Radioterapia Assistida por Computador / Irradiação Corporal Total / Radioisótopos de Cobalto / Imagens de Fantasmas / Radioterapia de Intensidade Modulada / Órgãos em Risco Limite: Humans Idioma: En Revista: J Appl Clin Med Phys Assunto da revista: BIOFISICA Ano de publicação: 2024 Tipo de documento: Article País de afiliação: Canadá País de publicação: Estados Unidos