Repeatability and Agreement of Central Vault for Implantable Collamer Lens Obtained by the Tomey OA-2000 Biometer and Spectralis OCT.
Wu, Hao; Wang, Zuocheng; Wang, Pengfei; Meng, Yifei; Wang, Zengying; Xue, Yuhong; Jiang, Bohua; Pan, Shuaixi; Yan, Zhipeng.
Afiliação
  • Wu H; Department of Ophthalmology Bengbu First People's Hospital, 229 Tushan Road, Bengbu 233000, China.
  • Wang Z; Department of Ophthalmology The Third Hospital of Hebei Medical University, 139 Ziqiang Road, Shijiazhuang 050000, China.
  • Wang P; School of Management Shijiazhuang Tiedao University, Shijiazhuang 050043, China.
  • Meng Y; Department of Ophthalmology The Third Hospital of Hebei Medical University, 139 Ziqiang Road, Shijiazhuang 050000, China.
  • Wang Z; Department of Ophthalmology The Third Hospital of Hebei Medical University, 139 Ziqiang Road, Shijiazhuang 050000, China.
  • Xue Y; Department of Ophthalmology The Third Hospital of Hebei Medical University, 139 Ziqiang Road, Shijiazhuang 050000, China.
  • Jiang B; Department of Ophthalmology The Third Hospital of Hebei Medical University, 139 Ziqiang Road, Shijiazhuang 050000, China.
  • Pan S; Department of Ophthalmology The Third Hospital of Hebei Medical University, 139 Ziqiang Road, Shijiazhuang 050000, China.
  • Yan Z; Department of Ophthalmology The Third Hospital of Hebei Medical University, 139 Ziqiang Road, Shijiazhuang 050000, China.
J Ophthalmol ; 2024: 3684626, 2024.
Article em En | MEDLINE | ID: mdl-38957378
ABSTRACT

Objective:

To assess repeatability and agreement of central vault for implantable collamer lens (ICL) measured by the Tomey OA-2000 biometry and Spectralis optical coherence tomography (OCT).

Methods:

In this prospective study, the central vault was measured by the Tomey OA-2000 biometer and Spectralis OCT in 84 eyes (43 patients) after ICL implantation at six month follow-up. Three consecutive scans were obtained by one experienced technician using Tomey OA-2000 and the Spectralis OCT in the same day. The coefficient of variation (CoV), intraclass correlation coefficient (ICC), within-subject standard deviation (Sw), and 2.77 Sw were calculated to assess the repeatability and reproducibility. The paired t-test and Bland-Altman plots were used to analyze the differences and agreements of central vault measured by two devices.

Results:

Repeatability of the central vault measured by Tomey OA-2000 biometer and Spectralis OCT showed that the CoV was 2.71% and 1.66%, respectively. The ICC for both devices was 0.996 and 0.999, respectively. The paired t-test showed that central vault measured by Tomey OA-2000 biometer was -7.25 ± 23.57 microns lower than that measured by Spectralis OCT (P = 0.006). The mean difference between measurements for Tomey OA-2000 and ASM-OCT with 95% limits of agreement (LoAs) was -38.94 to 53.44 µm.

Conclusion:

Both Tomey OA-2000 biometer and Spectralis OCT displayed good repeatability for the measurement of central vault of ICL. Good reliability and agreement were observed between Tomey OA-2000 biometer and Spectralis OCT. Both instruments are useful but not replaced each other for central vault measurements.