Your browser doesn't support javascript.
loading
Machine Learning Based Prediction of Post-operative Infrarenal Endograft Apposition for Abdominal Aortic Aneurysms.
van Veldhuizen, Willemina A; de Vries, Jean-Paul P M; Tuinstra, Annemarij; Zuidema, Roy; IJpma, Frank F A; Wolterink, Jelmer M; Schuurmann, Richte C L.
Afiliação
  • van Veldhuizen WA; Department of Surgery, Division of Vascular Surgery, University Medical Center Groningen, The Netherlands. Electronic address: w.a.van.veldhuizen@umcg.nl.
  • de Vries JPM; Department of Surgery, Division of Vascular Surgery, University Medical Center Groningen, The Netherlands.
  • Tuinstra A; Department of Surgery, Division of Vascular Surgery, University Medical Center Groningen, The Netherlands.
  • Zuidema R; Department of Surgery, Division of Vascular Surgery, University Medical Center Groningen, The Netherlands.
  • IJpma FFA; Department of Surgery, Division of Trauma Surgery, University Medical Center Groningen, The Netherlands.
  • Wolterink JM; Department of Applied Mathematics, Technical Medical Centre, University of Twente, Enschede, The Netherlands.
  • Schuurmann RCL; Department of Surgery, Division of Vascular Surgery, University Medical Center Groningen, The Netherlands; Multimodality Medical Imaging Group, Technical Medical Centre, University of Twente, Enschede, The Netherlands.
Article em En | MEDLINE | ID: mdl-38972630
ABSTRACT

OBJECTIVE:

Challenging infrarenal aortic neck characteristics have been associated with increased risk of a type Ia endoleak after endovascular aneurysm repair (EVAR). Short apposition (< 10 mm circumferential shortest apposition length [SAL]) on the first post-operative computerised tomography angiography (CTA) has been associated with type Ia endoleak. Therefore, this study aimed to develop a model to predict post-operative SAL in patients with an abdominal aortic aneurysm based on the pre-operative shape.

METHODS:

A statistical shape model was developed to obtain principal component scores. The dataset comprised patients treated with standard EVAR without complications (n = 93) enriched with patients with a late type Ia endoleak (n = 54). The infrarenal SAL was obtained from the first post-operative CTA and subsequently binarised (< 10 mm and ≥ 10 mm). The principal component scores that were statistically different between the SAL groups were used as input for five classification models, and evaluated by means of leave one out cross validation. Area under the receiver operating characteristics curves (AUC), accuracy, sensitivity, and specificity were determined for each classification model.

RESULTS:

Of the 147 patients, 24 patients had an infrarenal SAL < 10 mm and 123 patients had a SAL ≥ 10 mm. The gradient boosting model resulted in the highest AUC of 0.77. Using this model, 114 (78.0%) patients were correctly classified; sensitivity (< 10 mm apposition was correctly predicted) and specificity (≥ 10 mm apposition was correctly predicted) were 0.70 and 0.79, and were based on a threshold of 0.21, respectively.

CONCLUSION:

A model was developed to predict which patients undergoing EVAR will achieve sufficient graft apposition (≥ 10 mm) in the infrarenal aortic neck based on a statistical shape model of pre-operative CTA data. This model can help vascular specialists during the planning phase to accurately identify patients who are unlikely to achieve sufficient apposition after standard EVAR.