Multicenter CT Image-Based Anatomic Assessment of Patients with Aortoiliac Aneurysm Undergoing Endovascular Repair with Iliac Branch Devices.
Ann Vasc Surg
; 108: 484-497, 2024 Nov.
Article
em En
| MEDLINE
| ID: mdl-39009130
ABSTRACT
BACKGROUND:
The Global Iliac Branch Study (NCT05607277) is an international, multicenter, retrospective cohort study of anatomic predictors of adverse iliac events (AIEs) in aortoiliac aneurysms treated with iliac branch devices (IBDs).METHODS:
Patients with pre-IBD and post-IBD computed tomography imaging were included. We measured arterial diameters, stenosis, calcification, bifurcation angles, and tortuosity indices using a standardized, validated protocol. A composite of ipsilateral AIE was defined, a priori, as occlusion, type I or III endoleak, device constriction, or clinical event requiring reintervention. Paired t-test compared tortuosity indices and splay angles pretreatment and post-treatment for all IBDs and by device material (stainless steel and nitinol). Two-sample t-test compared anatomical changes from pretreatment to post-treatment by device material. Logistic regression assessed associations between AIE and anatomic measurements. Analysis was performed by IBD.RESULTS:
We analyzed 297 patients (286 males, 11 females) with 331 IBDs (227 stainless steel, 104 nitinol). Median clinical follow-up was 3.8 years. Iliac anatomy was significantly straightened with all IBD treatment, though stainless steel IBDs had a greater reduction in total iliac artery tortuosity index and aortic splay angle compared to nitinol IBDs (absolute reduction -0.20 [-0.22 to -0.18] vs. -0.09 [-0.12 to -0.06], P < 0.0001 and -19.6° [-22.4° to -16.9°] vs. -11.2° [-15.3° to -7.0°], P = 0.001, respectively). There were 54 AIEs in 44 IBDs in 42 patients (AIE in 13.3% of IBD systems), requiring 35 reinterventions (median time to event 41 days; median time to reintervention 153 days). There were 18 endoleaks, 29 occlusions, and 5 device constrictions. There were no strong associations between anatomic measurements and AIE overall, though internal iliac diameter was inversely associated with AIE in nitinol devices (nAIE, nitinol = 8).CONCLUSIONS:
Purpose-built IBDs effectively treat aortoiliac disease, including that with tortuous anatomy, with a high patency rate (91.5%) and low reintervention rate (9.1%) at 4 years. Anatomic predictors of AIE are limited.
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Desenho de Prótese
/
Prótese Vascular
/
Stents
/
Valor Preditivo dos Testes
/
Aneurisma Ilíaco
/
Implante de Prótese Vascular
/
Ligas
/
Procedimentos Endovasculares
/
Angiografia por Tomografia Computadorizada
Limite:
Aged
/
Aged80
/
Female
/
Humans
/
Male
/
Middle aged
País/Região como assunto:
America do norte
Idioma:
En
Revista:
Ann Vasc Surg
Assunto da revista:
ANGIOLOGIA
Ano de publicação:
2024
Tipo de documento:
Article
País de publicação:
Holanda