Early-stage Development of the CoRISMA Mechanical Circulatory Support (CMCS) System for Heart Failure Therapy.
Monreal, Gretel; Koenig, Steven C; Kelley, James F; Illg, Jessica J; Tamez, Daniel; Kelley, Mark S; Yetukuri, Varun; Cross, Daisy P; Theran, Michael E; Slaughter, Mark S.
Afiliação
  • Monreal G; Department of Cardiovascular and Thoracic Surgery, University of Louisville, 302 E. Muhammad Ali Blvd, Room 411, Louisville, KY, 40202, USA. gretel.monreal@louisville.edu.
  • Koenig SC; Department of Cardiovascular and Thoracic Surgery, University of Louisville, 302 E. Muhammad Ali Blvd, Room 411, Louisville, KY, 40202, USA.
  • Kelley JF; Department of Bioengineering, University of Louisville, Louisville, KY, USA.
  • Illg JJ; CoRISMA MCS Systems Inc, Hamden, CT, USA.
  • Tamez D; CoRISMA MCS Systems Inc, Hamden, CT, USA.
  • Kelley MS; CoRISMA MCS Systems Inc, Hamden, CT, USA.
  • Yetukuri V; CoRISMA MCS Systems Inc, Hamden, CT, USA.
  • Cross DP; CoRISMA MCS Systems Inc, Hamden, CT, USA.
  • Theran ME; CoRISMA MCS Systems Inc, Hamden, CT, USA.
  • Slaughter MS; CoRISMA MCS Systems Inc, Hamden, CT, USA.
Article em En | MEDLINE | ID: mdl-39037566
ABSTRACT

PURPOSE:

CoRISMA MCS Systems Inc (Hamden CT) is developing an innovative mechanical circulatory support system (CMCS) as a durable therapeutic option for heart failure (HF) patients. The CMCS system is comprised of an axial flow pump, non-contacting hydrodynamic bearings, and integrated DC motor designed to be fully implantable in a left atrial (LA) to aortic (Ao) configuration; this unloading strategy may be particularly beneficial for HF patients with preserved ejection fraction (HFpEF). The small (5.5 cm3), lightweight (20 g), and low power (5-7 W) device design should allow for a less invasive off-pump implant. We present early-stage engineering development and testing of the prototype CoRISMA pumps.

METHODS:

Computational fluid dynamics (CFD) modeling was performed to evaluate flow and shear in two impeller (3 blades, 0.5 mm thickness, 8.9 mm diameter, 0.15 mm gap, polished titanium) and diffusor (5 blades, polished titanium) candidate designs. Test apparatuses were custom built to expedite development of the impeller/diffuser designs and iteratively refine the CFD models. Two candidate impeller/diffusor designs were fabricated and tested in each of the two test apparatuses (n = 4 impeller/diffuser + test fixture configurations) in static mock flow loops (hydrodynamic H-Q curves, 3.5 cP glycerol solution at 37 °C), and in dynamic mock flow loops (hemodynamics, 3.5 cP glycerol solution at 37 °C) tuned to HF conditions (mean aortic pressure 50 mmHg, central venous pressure 15 mmHg, aortic flow 3.0 L/min, and heart rate 80 bpm).

RESULTS:

CFD predicted flows of 4.56 L/min and 4.82 L/min at 100 mmHg for impellers/diffusers 1 and 2, respectively. Impeller 2 required less torque to generate a 6% increase in fluidic flow, and the diffuser had a larger area of high pressure, indicative of lower friction, which likely contributed to the increased efficiency. Experimental testing for all four configurations in the static and dynamic mock loops met performance metrics as evidenced by generating 4.0-4.5 L/min flow against 70-76 mmHg pressure at 25,000 rpm and restoring hemodynamics in the dynamic mock flow loop (MAP = 80 mmHg, CVP = 0 mmHg, total flow = 5.5 L/min) from baseline simulated HF test conditions.

CONCLUSION:

These results demonstrate proof-of-concept of the early engineering design and performance of the prototype CoRISMA pumps. Engineering specifications, challenges observed, and proposed solutions for the next design iteration were identified for the continued development of an effective, reliable, and safe LA-to-Ao CMCS system for HF patients. Current design plans are underway for incorporating a wireless energy transfer system for communication and power, eliminating the need for and complications associated with an external driveline, to achieve a fully-implantable system.
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Revista: Cardiovasc Eng Technol / Cardiovasc. eng. technol. (Internet) / Cardiovascular engineering and technology (Internet) Ano de publicação: 2024 Tipo de documento: Article País de afiliação: Estados Unidos País de publicação: Estados Unidos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Revista: Cardiovasc Eng Technol / Cardiovasc. eng. technol. (Internet) / Cardiovascular engineering and technology (Internet) Ano de publicação: 2024 Tipo de documento: Article País de afiliação: Estados Unidos País de publicação: Estados Unidos