A rapid and validated GC-MS/MS method for simultaneous quantification of serum Myo- and D-chiro-inositol isomers.
Wang, Xiaofan; Lu, Lan; Peng, Huanqie; Li, Tanyao; Long, Qichen; Guan, Lihua; Xia, Xiaomeng; Li, Xianping; Wang, Min.
Afiliação
  • Wang X; Department of Laboratory Medicine, the Second Xiangya Hospital of Central South University, Changsha, Hunan 410011, China; Department of Clinical Laboratory, Taizhou Central Hospital (Taizhou University Hospital), Taizhou, Zhejiang Province, 318000, China.
  • Lu L; Physical and chemical laboratory of Hunan Provincial Center for Disease Control and Prevention, Changsha, Hunan 410005, China.
  • Peng H; Department of Laboratory Medicine, the Second Xiangya Hospital of Central South University, Changsha, Hunan 410011, China.
  • Li T; Physical and chemical laboratory of Hunan Provincial Center for Disease Control and Prevention, Changsha, Hunan 410005, China.
  • Long Q; Department of Laboratory Medicine, the Second Xiangya Hospital of Central South University, Changsha, Hunan 410011, China.
  • Guan L; Department of Laboratory Medicine, the Second Xiangya Hospital of Central South University, Changsha, Hunan 410011, China.
  • Xia X; Department of Obstetrics and Gynecology, the Second Xiangya Hospital of Central South University, Changsha, Hunan 410011, China.
  • Li X; Department of Laboratory Medicine, the Second Xiangya Hospital of Central South University, Changsha, Hunan 410011, China.
  • Wang M; Department of Laboratory Medicine, the Second Xiangya Hospital of Central South University, Changsha, Hunan 410011, China. Electronic address: wangmin0000@csu.edu.cn.
J Chromatogr A ; 1732: 465246, 2024 Sep 13.
Article em En | MEDLINE | ID: mdl-39128239
ABSTRACT

BACKGROUND:

Myo-inositol (MI) and D-chiro-inositol (DCI) are two paramount isomers of inositol, both vital in glucose and steroid metabolism. Deficits in MI, DCI or MI/DCI ratio are expressly concerned with several pathological process, whereas MI and DCI lack practical measurement for human specimen.

METHODS:

To quantify MI and DCI in serum samples simultaneously, a gas chromatography tandem mass spectrometry (GC-MS/MS) method was established. The process flow was optimized in ion source, derivative agent volume and reaction time. The performance characteristics were verified by commercial standards and clinical serums.

RESULTS:

This method was confirmed to be sensitive (LOD ≤ 30 ng/mL of MI, ≤3 ng/mL of DCI) and reproducible (RSD < 6 % for repeated analyses). Quantitative determinations performed good linearity within the measurement range of 0.500-10.00 and 0.005-0.500 µg/mL for MI and DCI respectively (R2 > 0.999). The recoveries of MI and DCI were 97.11-99.35 % and 107.82-113.09 %, respectively. This method was successfully applied to 114 clinical specimens. No significant matrix effect was observed in serum samples under current conditions.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Espectrometria de Massas em Tandem / Limite de Detecção / Inositol / Cromatografia Gasosa-Espectrometria de Massas Limite: Humans Idioma: En Revista: J Chromatogr A Ano de publicação: 2024 Tipo de documento: Article País de afiliação: China País de publicação: Holanda

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Espectrometria de Massas em Tandem / Limite de Detecção / Inositol / Cromatografia Gasosa-Espectrometria de Massas Limite: Humans Idioma: En Revista: J Chromatogr A Ano de publicação: 2024 Tipo de documento: Article País de afiliação: China País de publicação: Holanda