Performance of the MiniMed 780G system on mitigating menstrual cycle-dependent glycaemic variability.
Elhenawy, Yasmine Ibrahim; Abdel Kader, Mohamed S; Thabet, Rasha A.
Afiliação
  • Elhenawy YI; Paediatric and Adolescent Diabetes Unit (PADU), Department of Paediatrics, Faculty of Medicine, Ain Shams University, Cairo, Egypt.
  • Abdel Kader MS; Department of Paediatrics, Faculty of Medicine, Misr University of Science and Technology (MUST), Giza, Egypt.
  • Thabet RA; Paediatric and Adolescent Diabetes Unit (PADU), Department of Paediatrics, Faculty of Medicine, Ain Shams University, Cairo, Egypt.
Diabetes Obes Metab ; 2024 Sep 09.
Article em En | MEDLINE | ID: mdl-39248215
ABSTRACT

AIM:

To map the glycaemic variabilities and insulin requirements across different phases of the menstrual cycle and assess the efficacy and performance of the MiniMed 780G system on mitigating glycaemic variabilities during phases of the menstrual cycle. MATERIALS AND

METHODS:

A pilot study recruiting 15 adolescent and young adult females with type 1 diabetes was conducted. Only females with regular spontaneous menstruation were enrolled in the current study. Phases of each menstrual cycle were determined as either follicular phase or luteal phase. The study analysed continuous glucose monitoring metrics during two study periods the open loop period (OLP) and the advanced hybrid closed-loop (AHCL) period; each period lasted 3 consecutive months.

RESULTS:

During the OLP, the mean time in range (TIR) significantly decreased during the luteal phase compared with the follicular phase (65.13% ± 3.07% vs. 70.73% ± 2.05%) (P < .01). The mean time above range significantly increased from 21.07% ± 2.58% during the follicular phase to 24.87% ± 2.97% during the luteal phase (P < .01). After initiating the AHCL period, TIR was comparable during both phases of the menstrual cycle (P = .72), without increasing the time spent below 70 mg/dL (P > .05). Regarding insulin delivery during the AHCL period, the percentage of Auto basal and Auto correction delivered by the algorithm increased by 13.55% and 30.6%, respectively (P < .01), during the luteal phase.

CONCLUSIONS:

The fully automated adaptive algorithm of the MiniMed 780G system mitigated menstrual cycle-dependent glycaemic variability, successfully attaining the recommended glycaemic outcomes with a TIR greater than 70% throughout the entire menstrual cycle.