Rapid direct determination of lead in evaporated milk by anodic stripping voltammetry without sample pretreatment.
J Assoc Off Anal Chem
; 66(6): 1409-13, 1983 Nov.
Article
em En
| MEDLINE
| ID: mdl-6689015
ABSTRACT
The method presented describes the direct determination of lead in evaporated milk in which the milk ashing step prior to analysis is eliminated. Digital instrument readout units are microgram Pb/mL milk. Total analysis time after instrument calibration is less than 3 min per sample. Range of the method is 0.05-1.0 ppm lead in milk, and precision of the method expressed by relative standard deviation of duplicate pairs ranged from 30% at 0.1 micrograms/mL to 3% at 1.0 micrograms/mL of lead in milk. The method compares favorably with the AOAC official first action anodic stripping voltammetric method (25.074). In addition, the method appears to work equally well for skim evaporated milk, sweetened condensed milk, and nonfat powdered dry milk when the latter two are reconstituted with water according to product label instructions. Recovery and interference studies are presented.
Buscar no Google
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Análise de Alimentos
/
Chumbo
Limite:
Animals
Idioma:
En
Revista:
J Assoc Off Anal Chem
Ano de publicação:
1983
Tipo de documento:
Article