Early experience with the Jyros bileaflet valve. Is leaflet rotation clinically important?
Dihmis, W C; Wilde, P; Hutter, J A.
Afiliação
  • Dihmis WC; Department of Cardiac Surgery, Bristol Royal Infirmary, United Kingdom.
J Heart Valve Dis ; 3(2): 191-6, 1994 Mar.
Article em En | MEDLINE | ID: mdl-8012638
The design of the Jyros bileaflet prosthesis provides for a continuous, circumferential rotation of the hinge. Previous in vitro testing has confirmed the uninterrupted gyration of the leaflets, which reduces the risk of stasis within the hinge area, and as a consequence, possibly reduces the risk of thromboembolism. Twenty Jyros bileaflet valves were implanted in 19 patients between July 1991 and September 1992. There was one hospital death not related to the prosthesis. One valve had to be removed from the mitral position immediately after implantation because preservation of the posterior mitral leaflet led to malfunction of the prosthesis. Clinical and radiological follow up demonstrated that all surviving patients were clinically well, but nine of the remaining 18 valves (50%) showed no rotation of the leaflets.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Próteses Valvulares Cardíacas Tipo de estudo: Diagnostic_studies Limite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: J Heart Valve Dis Assunto da revista: CARDIOLOGIA Ano de publicação: 1994 Tipo de documento: Article País de afiliação: Reino Unido País de publicação: Reino Unido
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Próteses Valvulares Cardíacas Tipo de estudo: Diagnostic_studies Limite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: J Heart Valve Dis Assunto da revista: CARDIOLOGIA Ano de publicação: 1994 Tipo de documento: Article País de afiliação: Reino Unido País de publicação: Reino Unido