Dose-volume distributions: a new approach to dose-volume histograms in three-dimensional treatment planning.
Med Phys
; 21(1): 3-11, 1994 Jan.
Article
em En
| MEDLINE
| ID: mdl-8164585
A new approach to calculating and displaying dose-volume relationships in 3D radiation therapy is presented. We have developed a concept of a dose-volume distribution (DVD) and its corresponding differential dose-volume distribution (DDVD), based on organization of the data in the volume rather than in the dose domain. The new concepts make full use of the information that can be obtained from the dose calculation points and the sampling pattern and are designed to overcome shortcomings of the classical concepts of dose-volume histograms (DVH) and differential dose-volume histograms (DDVH). The new concepts can be applied to any number of dose calculation points, but they are especially advantageous when a small number of points is used. DVDs are particularly well suited to pseudo- and quasi-random sampling of dose distributions. We have developed an error analysis for DVDs and DDVDs in the case of pseudorandom sampling. We also describe an adaptive technique for minimizing the amount of data needed for purposes of display.
Buscar no Google
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Planejamento da Radioterapia Assistida por Computador
Limite:
Humans
Idioma:
En
Revista:
Med Phys
Ano de publicação:
1994
Tipo de documento:
Article
País de publicação:
Estados Unidos