Capillary electrophoresis for screening of adenylosuccinate lyase deficiency.
Electrophoresis
; 16(10): 1927-9, 1995 Oct.
Article
em En
| MEDLINE
| ID: mdl-8586067
We report a new screening method for adenylosuccinate lyase (ASase) deficiency using capillary electrophoresis (CE). This enzyme defect causes secondary autism and psychomotor retardation in early childhood. In all body fluids of these patients, two succinylpurine metabolites can be found that are normally not detectable: succinyladenosine and succinylaminoimidazole carboxamide (SAICA) riboside. A Beckman P/ACE 2050 capillary electrophoresis system was used with a 47.1 cm capillary, 75 microns ID, and the P/ACE Beckman UV absorbance detector. Untreated urine, injected for 1 s, was separated in a pH 8.63 borate buffer at 20 kV. The two succinylpurines (migration times 13.36 and 13.60 min) were detected at 254 nm only in urine of patients with ASase deficiency but not in control samples.
Buscar no Google
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Adenilossuccinato Liase
/
Eletroforese Capilar
Tipo de estudo:
Diagnostic_studies
/
Screening_studies
Limite:
Humans
Idioma:
En
Revista:
Electrophoresis
Ano de publicação:
1995
Tipo de documento:
Article
País de afiliação:
Alemanha
País de publicação:
Alemanha