Your browser doesn't support javascript.
loading
Geometric mismatch of the Gamma nail to the Chinese femur.
Leung, K S; Procter, P; Robioneck, B; Behrens, K.
Afiliação
  • Leung KS; Department of Orthopaedics and Traumatology, Chinese University of Hong Kong, Prince of Wales Hospital, Shatin.
Clin Orthop Relat Res ; (323): 42-8, 1996 Feb.
Article em En | MEDLINE | ID: mdl-8625605
The geometric mismatch of the femoral component of the Gamma nail to the Chinese femora resulted in intraoperative complications. To provide scientific data for modification of the implant, 28 pairs of normal Chinese femora were studied with the 3-dimensional reconstruction from the computed tomography scan data. Measurements were taken from the reconstituted drawings, and the anthropometric data were applied in the modification of the implant. This study is the first report on the application of Chinese anthropometric data on the design of a trauma implant. It is hoped that a larger scale of the study will provide a more comprehensive data base for wider application to orthopaedic implant design in the future.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Pinos Ortopédicos / Fraturas do Fêmur / Fêmur / Fixação Intramedular de Fraturas Limite: Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Clin Orthop Relat Res Ano de publicação: 1996 Tipo de documento: Article País de publicação: Estados Unidos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Pinos Ortopédicos / Fraturas do Fêmur / Fêmur / Fixação Intramedular de Fraturas Limite: Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Clin Orthop Relat Res Ano de publicação: 1996 Tipo de documento: Article País de publicação: Estados Unidos