Albumin standards and the measurement of serum albumin with bromcresol green. 1971.
Dumas, B T; Watson, W A; Biggs, H G.
Afiliação
  • Dumas BT; Department of Clinical Pathology, School of Medicine, University of Alabama, Birmingham 35233, USA.
Clin Chim Acta ; 258(1): 21-30, 1997 Feb 03.
Article em En | MEDLINE | ID: mdl-9049440
ABSTRACT
A rapid and reliable method for measuring serum albumin employing bromcresol green is described. The addition of albumin to a solution of bromcresol green in a 0.075 M succinate buffer pH 4.20 results in an increase in absorbance at 628 nm. The absorbance-concentration relationship is linear for samples containing up to 6 g/dl albumin. Bilirubin, moderate lipemia, and salicylate do not interfere with the analysis. The use of nonionic surfactant (Brij-35) reduces the absorbance of the blank, prevents turbidity and provides linearity. The results by this method agree very well with those obtained by electrophoresis and salt fractionation. The method is simple, it has excellent precision and the reagents are stable. A protein standard is introduced which can be employed for both the total serum proteins and albumin determinations.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Análise Química do Sangue / Albumina Sérica Limite: Humans Idioma: En Revista: Clin Chim Acta Ano de publicação: 1997 Tipo de documento: Article País de afiliação: Estados Unidos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Análise Química do Sangue / Albumina Sérica Limite: Humans Idioma: En Revista: Clin Chim Acta Ano de publicação: 1997 Tipo de documento: Article País de afiliação: Estados Unidos