Interventional MR imaging. Demonstration of the feasibility of the Overhauser marker enhancement (OMEN) technique.
Acta Radiol
; 38(1): 43-6, 1997 Jan.
Article
em En
| MEDLINE
| ID: mdl-9059400
PURPOSE: The poor localization facility of interventional instruments in MR imaging has been one of the major obstacles to the popularization of interventional MR imaging. It has been suggested that the Overhauser enhancement be used to generate markers of small size and high visibility. This article studies the feasibility of a localization marker based on this method. MATERIAL AND METHODS: A small Overhauser marker was constructed on the tip of a coaxial cable and comparative images were taken by a 0.23 T imager with and without electron spin irradiation. RESULTS: During irradiation an enhanced signal intensity from the marker was observed. The signal from the marker also exceeded the signal from a 0.25 mmol MnCl2 reference phantom. CONCLUSION: Its small size and high signal-to-noise ratio, together with immunity to most system nonlinearities and imaging errors, makes the Overhauser marker a promising localization method for the accurate positioning of interventional devices. The method may be applied at any field strength, and markers are visible in images obtained with any practical imaging sequence.