Your browser doesn't support javascript.
loading
Curing lights: changes in intensity output with use over time.
Poulos, J G; Styner, D L.
Afiliação
  • Poulos JG; Division of Prosthodontics and Restorative Sciences, New York University College of Dentistry, NY 10010, USA.
Gen Dent ; 45(1): 70-3, 1997.
Article em En | MEDLINE | ID: mdl-9171483
ABSTRACT
Intensity output of curing lights affects cure depth and time needed for complete polymerization of resin composites. In this study, changes in intensity output of Max Curing Lights (L.D. Caulk Co., Milford, DE) were compared over time. With a Demetron Model 100 Curing Radiometer (Demetron Research Corp., Danbury, CT), 201 curing lights (from 1 to 3 years old) were tested. The 1- and 2-year-old lights had mean outputs of 423 mW/cm2; the 3-year-old lights had a mean output of 376 mW/cm2. An analysis of variance (ANOVA) and post-hoc Scheffe test demonstrated a significant effect of use over time on intensity output of these lights.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Fontes de Energia Elétrica / Equipamentos Odontológicos Tipo de estudo: Prognostic_studies Idioma: En Revista: Gen Dent Ano de publicação: 1997 Tipo de documento: Article País de afiliação: Estados Unidos