Your browser doesn't support javascript.
loading
In vivo evaluation of an extracorporeal pediatric centrifugal blood pump.
Macha, M; Litwak, P; Yamazaki, K; Kameneva, M; Butler, K C; Thomas, D C; Taylor, L P; Griffith, B P; Borovetz, H S.
Afiliação
  • Macha M; McGowan Center, Department of Surgery, University of Pittsburgh Medical Center 15219, USA.
ASAIO J ; 43(4): 284-8, 1997.
Article em En | MEDLINE | ID: mdl-9242941
ABSTRACT
This paper summarizes the authors' in vivo experience in evaluating a miniature centrifugal blood pump designed for pediatric/neonatal ventricular support. Left ventricular bypass was accomplished in two adult sheep and five juvenile lambs (5.5-80.0 kg) via either central (left ventricle to carotid artery) or peripheral (jugular vein to carotid artery) cannulation. Animals were weaned from mechanical ventilation and continuously monitored. Hemodynamic parameters remained within a normal range over the duration of the bypass. Two of five lambs were electively killed at 8, and 76 hours; the remaining three lambs died from respiratory complications at 33, 44, and 156 hours. There were no mechanical complications, and blood seal integrity was confirmed beyond 6 days. The pump speed was maintained at 3,000-4,500 rpm with pump flow rates between 0.4-1.5 L/min. Average plasma free hemoglobin was below 20 mg/dl in the five lamb experiments. Renal, hepatic, and hematologic indices also remained within physiologic ranges. Histopathologic analyses of major organs revealed renal cortical infarctions in two of five lambs. Examination of the pump surfaces after explant indicated small areas of thrombus in the housing adjacent to the outflow ports in two experiments. These encouraging results support further testing and refinement of this miniature centrifugal pump.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Coração Auxiliar Tipo de estudo: Etiology_studies Limite: Animals / Child / Humans / Newborn Idioma: En Revista: ASAIO J Assunto da revista: TRANSPLANTE Ano de publicação: 1997 Tipo de documento: Article País de afiliação: Estados Unidos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Coração Auxiliar Tipo de estudo: Etiology_studies Limite: Animals / Child / Humans / Newborn Idioma: En Revista: ASAIO J Assunto da revista: TRANSPLANTE Ano de publicação: 1997 Tipo de documento: Article País de afiliação: Estados Unidos