Your browser doesn't support javascript.
loading
Carotid artery bifurcation: multiple-flip-angle, three-dimensional, time-of-flight MR angiographic technique.
Faro, S H; Mohamed, F B; Chen, C Y; Ortega, H V; Vinitski, S.
Afiliação
  • Faro SH; Department of Radiological Sciences, Allegheny University Hospitals-Medical College of Pennsylvania, Philadelphia 19129, USA.
Radiology ; 206(2): 555-9, 1998 Feb.
Article em En | MEDLINE | ID: mdl-9457212
ABSTRACT
Multiple-flip-angle, three-dimensional, time-of-flight magnetic resonance (MR) angiography was performed in vitro and in vivo in the carotid artery bifurcation. Composite MR angiographic images were created from multiple data sets. Compared with images obtained with a low (20 degrees) flip angle, composite images obtained with flip angles of 20 degrees and 60 degrees demonstrated improved image quality and statistically significant improvement in signal-to-noise ratio (P < .05).
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Processamento de Imagem Assistida por Computador / Artérias Carótidas / Angiografia por Ressonância Magnética Limite: Adult / Female / Humans / Male Idioma: En Revista: Radiology Ano de publicação: 1998 Tipo de documento: Article País de afiliação: Estados Unidos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Processamento de Imagem Assistida por Computador / Artérias Carótidas / Angiografia por Ressonância Magnética Limite: Adult / Female / Humans / Male Idioma: En Revista: Radiology Ano de publicação: 1998 Tipo de documento: Article País de afiliação: Estados Unidos